DC-2000C SERIES ULTRASONIC THICKNESS GAUGE

1. Switch off Mode

2. Gain setting
3. Language
4. Contrast
5. Default
6. Information
-Press
or
8.4.6.1 Switch off mode
-Press
into the menu
- Select Auto shut down after 1 Min. 3 Min. 5 Min.
- Press
to confirm.
8.4.6.2 Gain setting
In the user's measuring environment, both different materials and the same
material with different status will have different effects on the accurate and
stable measuring. So for different measured objects and different measuring
environment, users should adjust the work status of the instrument to meet more
measurements.
For many materials and measuring conditions, auto gain adjustment can be
used, but for some special measurement, adjusting the instrument's working
status is necessary. There are four different working modes: Auto, Low,
medium and high.
Auto: Match different probe and meet almost all the measuring requirements.
Low: Suitable for high scattering and small attenuation materials
Medium: Suitable for many measurements.
High: Suitable for high attenuation material
- Press
into the Menu
screen shows:
into the selected item, press
"6.Function"- "⑴Switch off mode"
" 6.Function"-"⑵Gain setting"
14
to confirm.
,