Purpose of use 10. Edition / 2019 – 06 The LED illumination set LI 900 was developed exclusively for Haag‑Streit slit lamps BQ 900, BP 900 and BM 900. The intensities of the slit and background illumination can be set via the illumination controls.
SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Contents Safety ................... 4 Decommissioning ..............9 Areas of application of the device ................4 Technical data ..............9 Ambient conditions ....................4 LI01 / LI02 plus .......................9 Shipment and unpacking ..................4 Power supply PS‑LED / PS‑LED HSM 901............10 Installation warnings ....................4 Dimensions ......................10 Operation, environment ..................4...
EN 60601-1 standard. The device is intended to use in professional health care facility environment, like • Only original Haag-Streit (HS) replacement parts may be used. doctor's practices, hospitals and optometrists and opticians premises, except near • The device must not be stacked or placed in close proximity to other of HF surgical equipment and in RF shielded rooms of ME‑systems for magnet‑...
Haag-Streit products must be used only for the purposes and in the manner des- • The illumination set LI 900 consists of an LED illumination unit, a power supply cribed in the documents distributed with the product. PS‑LED and an illumination control. A light conductor set is also required for the •...
Do not modify this equipment without authorization of the manufacturer. 8. Background illumination fixed Installation and repairs may only be performed by trained specialists. 9. Background illumination pivoting Contact your Haag-Streit representative for installation, repairs and modification work on the system. The contact details are available at Power supply PS-LED www.haag-streit.com.
Installation and repairs may only be performed by trained specialists. • Cables must not be trapped as this can cause a short circuit! Contact your Haag-Streit representative for installation, repairs and • Work on cables or device components, which are in contact with the modification work on the system.
ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Operating state Operation Green flashing Green flashing Standby mode briefly × briefly The light sources are switched on if the knob of the illumination control IC01 is in a position between ’1’ and ’10’. Normal mode Green Green...
To switch off the system NOTE! completely, the rocker switch must be moved to position 0 = ’OFF’. There is then • Observe the respective Haag-Streit instructions for use. (For further two‑pole isolation from the mains. information, please contact your Haag-Streit dealer). NOTE! •...
Installation and repairs may only be performed by trained specialists. USB illumination cable 2000 mm 1020940 Contact your Haag-Streit representative for installation, repairs and USB illumination cable 5000 mm 1020956 modification work on the system. The contact details are available at Cover T‑0 for recess in table...
ENGLISH B. Legal regulations • The illumination set LI 900 was designed t taking the standards EN 60601‑1, EN ISO 10939, EN 60601‑1‑2, and EN ISO 15004‑2 into account. • Compliance of the LI 900 illumination set with the Directive 93/42/EEC is con‑...
ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Information and manufacturer's declaration WARNING! • concerning electromagnetic compatibility Electrical medical devices and systems are subject to special EMC measures and must be installed in accordance with the EMC instruc- (EMC) tions contained in this accompanying document. General •...
SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Immunity (standard table 2) The information is based on the requirements of EN 60601‑1‑2:2007 (IEC 3 edition). Guidance and manufacturer's declaration – electromagnetic immunity This product is intended for use in the electromagnetic environment specifi ed below. The customer or the user of this product should assure that it is used in such an envi- ronment.
ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Immunity for non-life support devices (standard table 4) The information is based on the requirements of EN 60601‑1‑2:2007 (IEC 3 edition). Guidance and manufacturer's declaration – electromagnetic immunity This product is intended for use in the electromagnetic environment specifi ed below. The customer or the user of this product should assure that it is used in such an envi- ronment.
SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Safe distances for non-life support devices (standard table 6) The information is based on the requirements of EN 60601‑1‑2:2007 (IEC 3 edition). Recommended safe distances between portable and mobile HF communication devices and this device. This product is designed to be operated in an electromagnetic environment in which radiated HF interference is controlled.
Page 16
FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Should you have any further questions, please contact your Haag‑Streit representative at: http://www.haag-streit.com/contact/contact-your-distributor.html HAAG-STREIT AG Gartenstadtstrasse 10 3098 Koeniz, Switzerland Phone +41 31 978 01 11 +41 31 978 02 82 eMail info@haag-streit.com Internet www.haag-streit.com...