INSTRUCTION FOR USE
Photo slit lamp
BX 900
®
7. Edition / 2018 – 04
1
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland – HS Doc. No. 1500.7220449.04070 – 2018 – 04
01-IFU_BX900-7220449-04070_eng.indd 1
03.05.2018 09:47:11

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BX 900 and is the answer not in the manual?

Questions and answers

Summary of Contents for Haag-Streit BX 900

  • Page 1 PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH INSTRUCTION FOR USE Photo slit lamp BX 900 ® 7. Edition / 2018 – 04 © HAAG‑STREIT AG, 3098 Koeniz, Switzerland – HS Doc. No. 1500.7220449.04070 – 2018 – 04 01-IFU_BX900-7220449-04070_eng.indd 1 03.05.2018 09:47:11...
  • Page 2: Table Of Contents

    Introduction Warranty and product liability ..................5 Description of symbols ....................5 Thank you for choosing a HAAG-STREIT device. Provided you comply carefully Introduction ................6 with the regulations in this instructions for use, we can guarantee the reliable and Overview ........................6 unproblematic use of our product.
  • Page 3 SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Operation ................10 Appendix ................15 Setting the eyepieces ....................10 Accessories / spare parts ..................15 Preparing the patient ....................10 Legal regulations ..............16 Operating the instrument ..................10 5.4 Periphery or background illumination (flash)............11 C. Classification ..............16 Photo diaphragm pre-selection ................11 Disposal ................
  • Page 4: Safety

    ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Safety tative of the transport company. Make a report of any damaged parts. This report must be signed by you and by the representative of the transport company. DANGER! • Leave the device in the packaging for a few hours before unpacking it (conden- Failure to comply with these instructions may result in material damage sation).
  • Page 5: Light Toxicity

    Warranty and product liability number of pulses to 10. • Haag-Streit products must be used only for the purposes and in the manner des- NOTE! cribed in the documents distributed with the product. • The radiation times and number of total pulses of all light sources are •...
  • Page 6: Introduction

    40. Camera cable 12. Weight compensation screws 41. Ocular with double cross hairs After turning off the FU 01 flash 13. Trade mark BX 900 42. Mirror and diaphragm component unit, wait at least 15 seconds 14. Slit length / diaphragm scale 43. Cable trigger key for mirror/ before the flash cable is re- 15.
  • Page 7 SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Diffusor 1007432 © HAAG‑STREIT AG, 3098 Koeniz, Switzerland – HS Doc. No. 1500.7220449.04070 – 2018 – 04 01-IFU_BX900-7220449-04070_eng.indd 7 03.05.2018 09:47:17...
  • Page 8: Lens Barrel

    55. Optical light guide connection for slit and periphery/background illumination camera models and makes the necessary adjustments. The photo slit lamp is only 56. Filter/diaphragm wheel for periphery/background illumination (flash) This wheel guaranteed to function properly if cameras recommended by HAAG-STREIT are is used to control the intensity of the flash periphery/background illumination used. Please consult your local partner for information on the supported cameras.
  • Page 9: Device Assembly / Installation

    Do not modify this equipment without authorization of the manufacturer. Installation and repairs may only be performed by trained specialists. 78. Lens barrel Contact your HAAG-STREIT representative for installation, repairs and 79. Mirror and diaphragm component modification work on the system. The contact details are available at 80.
  • Page 10: Commissioning

    • The device must be switched off after every use. Otherwise there is a Setting the eyepieces risk of overheating when a protective dust cover is employed. • The slit lamp BX 900 may only be operated with the HS Flash Unit NOTE! FU 01 The eyepieces must be individually set prior to the first examination in •...
  • Page 11: Periphery Or Background Illumination (Flash)

    SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH turning the illumination facility on the handle (15) (locking points at 45°, 90° and 86. Rotating knob for diaphragm actuation 135°; stops at 0° and 180°; scale in 5° increments). Turning one of the two knobs on the right and left allows you to close the dia- • To ensure that unimpeded binocular fundus examination is also possible at lateral phragms to the preset aperture during observation in order to get a preview of the angles of between 3°...
  • Page 12: Microscope And Eyepiece

    NOTE! Capture frequency ‘high’ All red LEDs are flashing (electronic defect) Captures 1 to 10 in a 1-second cycle • Contact your HAAG-STREIT representative! Captures 11 to 15 in a 2-second cycle Decommissioning Captures 16 to 20 in a 3-second cycle Captures 21 to 30 in a 5-second cycle Throw the rocker switch on the PS-LED power supply and the main switch on the FU 01 flash unit 01 I = ’OFF’. The green lights on the rocker switches will stop illu-...
  • Page 13: Stereo Microscope

    SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Flash illumination Flash unit FU 01 Spectral range slit illumination 400 to 750 nm Spectral range background 400 to 750 nm Spectral range slit flash illumination 400 to 750nm illumination Spectral range background flash illumination 400 to 750nm Slit image width 0 –...
  • Page 14: Lens Barrel

    Installation and repairs may only be performed by trained specialists. • Slide the coupling slide back downward until it Contact your HAAG-STREIT representative for installation, repairs and stops in place. modification work on the system. The contact details are available at NOTE! www.haag-streit.com.
  • Page 15: Servicing

    Contact your HAAG-STREIT representative for installation, repairs and Cleaning and disinfection modification work on the system. The contact details are available at The Haag-Streit slit lamps and their accessories can, if required, be carefully wiped www.haag-streit.com. down with ready-for-use disposable 70% ethanol disinfectant wipes. Surface-friend- •...
  • Page 16: Legal Regulations

    NEDERLANDS SVENSKA Legal regulations NOTE! • The BX 900 photo slit lamp was developed and designed taking the EN 60601-1, An asterisk (*) indicates that you should contact your HAAG-STREIT EN ISO 10939 and EN ISO 15004-2 standards into account. representative for further information. • The EN 60601-1 standard must be observed when using different medical and/or Two asterisks (**) indicate a need to refer to the separate instructions non-medical electrical devices in combination.
  • Page 17: Information And Manufacturer's Declaration Concerning Electromagnetic Compatibility (Emc)

    SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Information and manufacturer's declaration WARNING! • concerning electromagnetic compatibility Electrical medical devices and systems are subject to special EMC measures and must be installed in accordance with the EMC instruc- (EMC) tions contained in this accompanying document. •...
  • Page 18: Immunity (Standard Table 2)

    ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Immunity (standard table 2) The information is based on the requirements of EN 60601-1-2:2007 (IEC 3 edition). Guidance and manufacturer's declaration – electromagnetic immunity This product is intended for use in the electromagnetic environment specifi ed below. The customer or the user of this product should assure that it is used in such an envi- ronment.
  • Page 19: Immunity For Non-Life Support Devices (Standard Table 4)

    SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Immunity for non-life support devices (standard table 4) The information is based on the requirements of EN 60601-1-2:2007 (IEC 3 edition). Guidance and manufacturer's declaration – electromagnetic immunity This product is intended for use in the electromagnetic environment specifi ed below. The customer or the user of this product should assure that it is used in such an envi- ronment.
  • Page 20: Safe Distances For Non-Life Support Devices (Standard Table 6)

    ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Safe distances for non-life support devices (standard table 6) The information is based on the requirements of EN 60601-1-2:2007 (IEC 3 edition). Recommended safe distances between portable and mobile HF communication devices and this device. This product is designed to be operated in an electromagnetic environment in which radiated HF interference is controlled.
  • Page 21 SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH © HAAG‑STREIT AG, 3098 Koeniz, Switzerland – HS Doc. No. 1500.7220449.04070 – 2018 – 04 01-IFU_BX900-7220449-04070_eng.indd 21 03.05.2018 09:47:30...
  • Page 22 ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Should you have any further questions, please contact your HAAG-STREIT dealer at: http://www.haag-streit.com/contact/contact-your-distributor.html HAAG-STREIT AG Gartenstadtstrasse 10 3098 Koeniz, Switzerland Phone +41 31 978 01 11 +41 31 978 02 82 eMail info@haag-streit.com...