AP-300/AP-600
HMI Quick User Guide

Advertisement

loading

Summary of Contents for Nordson AP-300

  • Page 1 AP-300/AP-600 HMI Quick User Guide...
  • Page 2 START THE PLASMA UNIT 1) Start the vacuum pump (vacuum pump could be a different type) Start the pump 15 minutes prior using the plasma system. 2) Check by rotating counter clockwise the EMO button is released 3) Press the green On button to power-up the Plasma system.
  • Page 3: How To Login

    HOW TO LOGIN After the system has been powered and initialized press (1) to goto the Main Screen. Press the Logout indicator box (2) The Password Pop-up Screen appears. Type in the password for the desired user level and click ENT (3) to submit de code System Default passwords per level below;...
  • Page 4 CREATE/DEVELOP A RECIPE (1/3) • After power up and initializing stage press (1) to goto the Main overview screen. • Press Recipe Name(2) to access the recipes screen • Select an empty memory storage location or a recipe to overwrite(3) •...
  • Page 5 CREATE/DEVELOP A RECIPE (2/3) • Make sure the system is in write Reminder, where to find the mode(1) manual mode alarms switch: • Make sure the system is in manual mode. Stop (7) should have a green cadre prior the Manual (2) can be activated in case it is off.
  • Page 6 CREATE/DEVELOP A RECIPE (3/3) Parameters Info: FWD: Forwarded Power, Power applied to the electrode(s) REF: Reflected Power, should be ≤ 7 Watts. When tuning in Auto mode automatically reduced to minimum value,ideal 0 Watt. TIME: RF on time while reflected power ≤...
  • Page 7 SHUTDOWN THE PLASMA UNIT WITH CHAMBER UNDER VACUUM 1) From the Main screen press Manual 2) In the Manual screen press Manual when the buttons cadre is not green 3) Press Vacuum to pull the chamber vacuum. When the pressure reading is <100mT press the EMO button(4) to shutdown the plasma unit 5) Switch the vacuum pump off.