Invacare SoftCloud Ace User Manual page 28

Alternating matress systems
Hide thumbs Also See for SoftCloud Ace:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
SoftCloud® Ace, SoftCloud® Pro
Warning/Fault
Low pressure indicator LED
lights up
+ audible warning
Alternation Failure indicator
LED lights up
+ audible warning
Power loss indicator LED
lights up
+ audible warning
Service Indicator LED lights
up
28
Cause
Pressure too low
Alternating Mode Failure (no
alternation)
Power failure
Service needed
Solution
1. Mute the audible warning and turn off power.
2. Check that the handle is intact, ensuring all four sealing connectors are
firmly attached to control unit and air hoses. Check the CPR Tag is
attached and all four sealing connectors are firmly secured.
3. Check all air hoses along the inside of the mattress – each should be
firmly connected. Check each air cell is securely attached to its connecting
air pipe.
4. Check all cells, pipes and hoses for any air leakage.
5. Check that the air filter shroud is correctly secured and the air filter
is clean.
6. Turn on Power.
1. Mute the audible warning and turn off power.
2. Disconnect the air hoses to reduce pressure – reconnect when pressure
has decreased.
3. Restart the system by switching the power button off and on. The system
automatically starts in Static Mode and will switch to Alternating Mode
when operational pressure is reached.
4. Reset the personal settings.
1. Mute the audible warning.
2. Check that the power cable is firmly plugged into the mains outlet and the
control unit. Check that the mains power is switched on.
3. Check both control unit fuses (1 amp slow blow fuse) – fuse can be
released using a screwdriver to push and turn.
If power is restored within 20 minutes of failure, the system will run an initial
start up sequence before returning to the last settings.
1. Contact an Invacare provider or Invacare representative.
There is no risk for continued use even if the Service Indicator LED
lights up.