Testing - LO90 Models
Doc 0223520 rev N
Locate the pin for the bat door. Push it out and remove the bat door. Add the correct prog adapt (pediatric or non pediatric) and fix it with the pin.
Insert batt pill size 675
Gently close the batt door and connect the CS44 prog cable. Place the fixture in the test chamber. Insert the hook into the 2cc coupler tube and
place it as shown (make sure the HI does not touch neither the fixture nor the ATC4000 sound chamber). Connect the extension cable to the
DSA6000 as also shown on page "Testing - LO85 Models". Select the correct test tpix and run the test making sure to follow instructions coming
up on the screen