ReSound ENZO QUATTRO Series Technical Manual page 36

Rs london platform
Table of Contents

Advertisement

HI Testing - LO85 Models
Use of test fixture is mandatory during DSA6000 final HI testing. The HI fixture/jig is to be placed in the black test frame. Place the HI in the fixture
according to pic and place it all in the test chamber and select the correct test tpix and make sure to follow instructions coming up on the screen.
During the test programme execution the IAD calibration will be performed at the appropriate time. Another benefit from the fixture is that the HI
is in a well defined position so the telecoil test can be performed without any interaction from the operator
2cc coupler
2cc coupler insert BTE
p/n 30-4838800
p/n 40B4609601
Not subject to issue control when printed
Coupl Mic,TM-12,measured
p/n 18220700
20 mm
The test tubing is normal 20 mm effective
lenght as for all BTE testing
Doc 0223520 rev N
Page 36 of 39