Staining Of Specimens - Thermo Scientific Microm HMS 740 Instruction Manual

Robot stainer
Table of Contents

Advertisement

5-2

STAINING OF SPECIMENS

Turn on the instrument.
Note:
Now the initialization phase is carried
out. The
transportation arm moves
into its end position to a stop. After this
phase, the main screen is shown on the screen.
Open the drawer.
Remove the cover from the entry station.
Insert the basket into the entry station.
Close the drawer.
Select a protocol either via "Start Protocol" on
the touch screen or via a quick-start button.
Note:
If the protocol is to be started in the
first step, press the quick-start button
twice.
To avoid position conflicts, equivalent baths
should be defined.
To better interlace protocols flexible times can
be entered.
Note:
The flexible time of a station is added
to the programmed minimum duration
time and varies according to the
number and requirements of parallel running
protocols.
Warning:
After the daily routine work has been
completed, the water flow to the
instrument must be turned off.
After having finished work, turn off the
instrument in a controlled way.
54
Thermo Scientific Robot Stainer Microm HMS 740
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387801 - English