UniData Communication Systems Starflow QSD Manual page 26

Table of Contents

Advertisement

Manual – Starflow QSD Model 6527B and 6537A
In the diagram above there are settings associated with the "CW (Continuous Wave) Velocity
Settings. These settings control the way velocity readings are returned by the QSD instrument.
Zero Result on RSSI fails - This will return a zero velocity reading if the RSSI value Is below a certain
point. This value is set in the Min RSS(Signal) box.
Zero Result on SPREAD fail - This will return a zero velocity if the SPREAD value is above a certain
point. This value is set in the Max Signal Spread box.
Min Depth (mm) - This will return a zero velocity reading if the depth falls below a certain value.
Unidata Manual - 6527B and 6537A Starflow QSD User Manual 17 01 2020.docx
Page 24

Advertisement

Table of Contents
loading

This manual is also suitable for:

6527b6537a

Table of Contents