4.1.4 Starflow QSD - SET/RESET Option Settings (aX5)
To alter the operation of QSD, there are 32 option settings numbered 0 to 31. These affect QSD's
performance so must be changed "wisely". The aX5 eXtended command is used and several options
adjustments (set and/or reset) can be made using a single aX5 command. A SET option is a
<+value> and RESET a <-value>. The <value> is the option number 0..31. Undefined options MUST
NOT be changed.
Option No.
SET Meaning
<value>
< + value>
04
Enable Pressure Depth
05
Zero P.Depth on Fail
06
Enable Barometric Ref
07
Show –ve P.Depth
08
Enable Electrical Conductivity (EC) Disable Electrical Conductivity (EC=0)
16
Enable Depth Measurement
17
Depth = 0 if < Min Operating Level
18
Enable Depth Drive 1
19
Enable Depth Drive 2
20
Enable Depth Drive 3
21
Enable Depth Drive 4
22
Use MEDIAN value for Depth
24
Enable CW Velocity Measurement
25
CW Velocity = 0 if < MIN RSSI
26
Enable -ve CW Velocity
27
Enable SOS Temp Corr.
31
Velocity = 0 if > MAX SPREAD
Examples:
Command
Response
aX03!
a3<
CR><LF>
aX1!
a+9+0+1+2+3+4+5+72+12
+13+0+0+0+0+0+0+0
aX1+6!
a+6+0+1+2+3+4+5+72+12
+13+0+0+0+0+0+0+0
aX2!
a +1150+40+4+5000
aX2+1200!
a+1200+25+1+300
aX4!
a+3+0+3+0+3+0+3+0+3+0
aX4+3+2!
a+3+2+3+0+3+0+3+0+3+0
a+1+16+17+18+19+20+22+24+27+30
aX5!
aX5-18+21
a+1+16+17+19+20+21+22+24+27+30
aX7-33!
a-33<CR><LF>
aX#!
a+nnnnn<CR><LF>
Notes: **
The Sensor address is an ASCII value so DOES NOT use a ± sign.
***
Only SET Options are displayed. Change order is not important.
****
Except for aX5, the order of changed parameters IS important. You MUST
re-enter all parameters occurring prior to the parameter you wish to change.
Unidata Manual - 6527B and 6537A Starflow QSD User Manual 17 01 2020.docx
Manual – Starflow QSD Model 6527B and 6537A
<CR><LF>
>
<CR><LF>
<CR><LF>
<CR><LF>
<CR><LF>
<CR><LF>
<CR><LF>
RESET Meaning
< - value>
Disable Pressure Depth
Retain last P.Depth
Disable Barometric Ref.
P.Depth +ve only
Disable Depth Measurement (Depth = 0)
Retain last measurement
Disable Depth Drive 1
Disable Depth Drive 2
Disable Depth Drive 3
Disable Depth Drive 4
Use MAXIMUM value for Depth
Disable CW Velocity (Velocity = 0)
Retain last measurement
+ve Velocity regardless of direction
Don't correct for SOS
Retain last measurement
Meaning
Sensor address was "a" and now "3" **
aD0/R0! will return 9 measurements
from Channel Nos. listed
Now will return 6 measurements (max 16)
st
th
from 1
to 6
Low battery limit is 11.50 Volts
Depth>40mm, RSSI>4,Spread>5000
Low battery limit changed to 12.00 Volts
Examine Outlier filter & Average setting
Change Moving Average from 0 to 2
Default SET Options***
Disable drive level 1, enable 4
Pressure Depth Offset = –33 mm of water
Restore Factory Defaults nnnnn=QSD Serial#
Channel Nos. listed
Page 11