Required Utilities

The following facility utilities are required to operate the Mini2. The provisions for these facilities
are the responsibility of the customer and are not included in the scope of supply for this
machine.
Physical Location The machine should be installed in a location that is accessible
Dimensions Coating Unit
Electrical Power Coating Unit
Compressed Air 80 psig, 10 cfm. The connection on the machine is a 1/4" quick
Deionized Water Consumption of 7.6 L/hr at periodic rates of 0.3 L/min of
Water Drain The machine is supplied with a drain reservoir that needs to be
www.ultraoptics.com
for easy disconnect of power and has good ventilation with an
air exchange of 4-5 times per hour. It should be located in a low
traffic area to minimize possible dust contamination. While it is
not necessary to install the Mini2 in a clean room, it is not a
good idea to place it near a lens edger either.
21" D x 25" W x 36" H
Cure Unit
25" D x 16" W x 13" H
115 VAC, 5 amp circuit required. Machine is supplied
with a grounded three-prong receptacle, OR
One single phase 220 VAC/5A circuit.
Cure Unit
115 VAC, 15 amp circuit required. Machine is supplied
with a grounded three-prong receptacle, OR
One single phase 220 VAC/12A circuit.
connect
deionized water flow rate with ≥ 1 megaohm resistivity. The
machine is supplied with a reservoir that needs to be
periodically filled by the user. The pH value should be greater
than 4.9 and less than 9.1.
periodically dumped by the user.