4. Unscrew the bottom fitting and place filter above your clean bottle and release the bleed
valve.
a. This will allow the coating to drain out of the filter.
5. Discard the filter.
6. Use Teflon tape on your new filter where the fittings will be screwed into.
7. Pour acetone into the coating bowl, just as if adding coating.
8. With the top fitting from the coating filter into a waste container, turn on the coating pump
and allow the acetone to flush out until empty.
9. Repeat flushing with acetone one more time to ensure any and all contaminates have
been removed.
10. Allow ten minutes for the reservoir to dry.
11. Attach the new coating filter and ensure the fittings are tight and do not leak.
12. Pour at least 1-1/2 bottles of new coating into the system and run coating prime for
approximately seven minutes to re-prime and to make sure no air and debris are trapped
in the coating filter.

Air Regulator Adjustment

1. Locate regulator on back panel
2. If an adjustment needs to be made, pull down on the regulator adjustment cap and turn
until the air gauge reads 80 PSI.
3. Push adjustment cap back up to lock in place.
www.ultraoptics.com
page 14