Section 5 Startup; Section 6 User Navigation; Section 7 Operation; System Configuration - Hach 9582sc User Manual

Hide thumbs Also See for 9582sc:
Table of Contents

Advertisement

Section 5 Startup

Make sure that the flow rate and pressure do not exceed the values in
1. Open the valve on the sample line to let sample flow through the analyzer.
2. Turn the knob on the flow meter to set the flow rate.
3. Examine the plumbing for leaks and stop any leaks if found.
4. Apply power to the controller.
5. Make the applicable menu selections when the controller starts.

Section 6 User navigation

Refer to the controller documentation for keypad description and navigation information.
Push the RIGHT arrow key on the controller multiple times to show more information on the home
screen and to show a graphical display.

Section 7 Operation

7.1 System configuration

Refer to the controller documentation for system configuration, general controller settings and setup
for outputs and communications.

7.2 Configure the sensor

Use the CONFIGURE menu to enter identification information for the sensor and to change options
for data handling and storage.
1. Push the menu key, select SENSOR SETUP>CONFIGURE.
2. Use the arrow keys to select an option and push enter. To enter numbers, characters or
punctuation, push and hold the up or down arrow keys. Push the right arrow key to advance to
the next space.
Option
EDIT NAME
SENSOR S/N
MEAS UNITS
PRESSURE UNITS Sets the units for atmospheric pressure—Select the unit from the list available.
TEMP UNITS
FILTER
LOG SETUP
RESET DEFAULTS Sets the configuration menu to the default settings. All sensor information is lost.
10 English
Description
Changes the name that corresponds to the sensor on the top of the measure screen.
The name is limited to 16 characters in any combination of letters, numbers, spaces or
punctuation.
Allows the user to enter the serial number of the sensor, limited to 16 characters in any
combination of letters, numbers, spaces or punctuation.
Changes the measurement units—Select the unit from the list available.
Sets the temperature units to °C (default) or °F
Sets a time constant to increase signal stability. The time constant calculates the
average value during a specified time—0 (no effect, default) to 60 seconds (average of
signal value for 60 seconds). The filter increases the time for the sensor signal to
respond to actual changes in the process.
Sets the time interval for data storage in the data log—5, 30 seconds, 1, 2, 5, 10,
15 (default), 30, 60 minutes.
Specifications
on page 3.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Polymetron 9 82 series

Table of Contents