Signal circuit
Basic specification, Position 3
(Connect. Power Supply;
Connect. Output) = B, E, J, K
Basic specification, Position 3
(Connect. Power Supply;
Connect. Output) = C, D, G, H
Terminal compartment B
Intrinsic safety (Ex ia)
 10
Signal circuit
4...20 mA/ HART
U
1)
(active)
I
o
P
+ -
R
C
L
Characteristic
curve: linear
U
I
i
P
PT100
U
PT100
I
o
P
R
C
L
Characteristic
curve: linear
2)
Cascade out
U
I
o
P
- +
R
C
L
Characteristic
curve: linear
3)
Cascade in
U
I
i
P
+ -
C
L
not connected
+ -
4...20 mA/HART (active)
U
= 253 V
m
AC
The detector ensures galvanic isolation up to a maximum of
+ -
253 V
between the signal circuit and any other circuit.
AC
OFF
ON
SIM WP
PT100
9 10 1112
IN CASCADE OUT
+ –
+ –
5 6
3 4
Ex ia IIC
= 21.2 V
C
= 150 nF
o
o
= 92 mA
at
= 479 mW
L
= 0.15 mH
o
o
= 235 Ω
C
= 115 nF
i
o
= 13.4 nF
at
i
= 0
L
= 1 mH
i
o
= 30 V
i
= 13 mA
= 390 mW
i
= 8.4 V
C
= 1 200 nF
o
o
= 8.3 mA
at
= 17.5 mW
L
= 1 mH
o
o
= 1 012 Ω
C
= 1 800 nF
i
o
= 0
at
i
= 0
L
= 0.15 mH
i
o
= 8.4 V
C
= 5.1 µF
o
o
= 19.2 mA
L
= 69 mH
o
= 40.3 mW
o
= 439 Ω
i
= 5.3 nF
i
= 67 µH
i
= 8.4 V
i
= 19.2 mA
= 40.3 mW
i
= 0
i
= 67 µH
i
– +
7 8
Ex ia IIB
Ex ib IIC
C
= 886 nF
C
= 169 nF
o
o
at
L
= 4 mH
o
L
= 0.15 mH
o
C
= 686 nF
o
at
L
= 1 mH
o
C
= 6 µF
C
= 5.2 µF
o
o
at
L
= 400 mH
o
L
= 1 mH
o
C
= 5.2 µF
o
at
L
= 2 mH
o
C
= 42 µF
C
= 5.1 µF
o
o
L
= 199 mH
L
= 69 mH
o
o
XA00303F-G
A0031764
Ex ib IIB
C
= 1.2 µF
o
L
= 15 mH
o
C
= 43 µF
o
L
= 400 mH
o
C
= 42 µF
o
L
= 199 mH
o