Endress+Hauser Gammapilot M FMG60 Safety Instructions page 20

Hide thumbs Also See for Gammapilot M FMG60:
Table of Contents

Advertisement

Available languages

Available languages

XA00303F-G
20
Circuit 4...20 mA/HART (active) in type of protection Increased safety (Ex eb) or
Flameproof enclosure (Ex db)
Basic specification, Position 3 (Connect. Power Supply; Connect. Output) = C, D, G, H
A
4
3
B
2
1
 8
A
Zone 1 or Zone 2
B
Zone 0, Zone 1 or Zone 2
1
Local potential equalization line
2
PT100; approved Ex ia equipment
3
Gammapilot with NaJ crystal scintillator, PVT plastic scintillator
4
[Ex ia] circuit
5
Terminal compartment A (Ex db or Ex eb)
6
Terminal compartment B (Ex i)
7
[Ex ia] circuit; Cascade in, out
8
[Ex ia] circuit
9
Communication: 4...20 mA/HART active
10
Power supply
11
E.g. transmitter
12
Remote display FHX40
13
Gammapilot with NaJ crystal scintillator, PVT plastic scintillator
• To maintain the ingress protection of the housing IP66/67, install the housing cover, cable glands
and blind plugs correctly.
• Cable glands as well as sealing plugs of the terminal compartment A must not be exchanged with
those of the terminal compartment B.
• Do not open the terminal compartment A when energized.
• In an explosive atmosphere: Minimum waiting time before opening the terminal compartment A
after switching off the power supply: 3 minutes.
• Before operation:
– Screw in the cover all the way.
– Tighten the securing clamp on the cover.
• Continuous service temperature of the connecting cable: ≥ T
5
6
5
Gammapilot M FMG60
7
8
9
10
11
11
12
6
13
+20 K.
a
Endress+Hauser
A0031755