Go through the "Switch ON procedure" only 1-2 hours before using it!
-
Turn on the N
flow on the "wall
2
valve).
-
Turn on the "red box" (behind the box).
-
Adjust the N
flow on the little black wheel to 12-20l/min.
2
-
Turn on the cooling water
N
flow is 12-20l/min and the cooling water is on? If yes go ahead:
2
-
Tun on lamps of the power supply: POWER1, then POWER 2
-
Switch on the temperature controller (PTC-348MI) and press "Start".
-
Start the computer and start the software on the desktop.

Sample Preparation:

1) Prepare your sample in a volumetric flask, don't fill it into a cuvette.
Check on the UV/vis spectrometer, wheter you are in the right concentration range.
- Strong chromophores: c ≈ 10
- Weak chromophores: c ≈ 10
- GENERAL RULE: your absorbance band of interest should be at 0.1-0.8 E, ideally at around
0.2-0.4 E.
2) Fill your cuvette with pure solvent.
Start your measurement: (after the machine warmed up for 1-2 hours)
-
Choose "Setup", then setup some parameters which you find in the submenus.
-
Insert your cuvette with pure solvent into the machine and record a Baseline. Save it.
-
Remove cuvette, empty it, blow it out with N
machine.
-
Measure your sample.
-
Measure as many samples as your want, all will be automatically be corrected your baseline
file.

Turn off the machine:

-
close the Specta Manager software shut down the computer
-
turn off the PTC-348WI turn off Power 2 switch
-
turn off Power 1 switch
-
purge instrument with N
column". Only use the black wheel (not the pressure-reducing
M. If necessary, start diluting it after a first quick measurement.
-5
M. If necessary, start diluting it after a first quick measurement.
-4
for 5 minutes after shutdown before turning off the nitrogen flow.
2
and fill your sample solution. Put it into the
2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the J-715 and is the answer not in the manual?

Questions and answers

Summary of Contents for Jasco J-715

  • Page 1 Manual for CD spectrometer Subscribe to the outlook calendar (see below) Go through the “Switch ON procedure” only 1-2 hours before using it! Turn on the N flow on the “wall column”. Only use the black wheel (not the pressure-reducing valve).
  • Page 2 on/off N flow control N cooling water PTC-348WI power supply...
  • Page 3 Reservation for CD J-710 DCHAB-user only Outlook-calendar...

Table of Contents