3.3 More about Dark Calibration

If the Sensor Cover is covering the Sensor while you power
on, a Dark Calibration is performed – otherwise Dark
Calibration is not performed.
It is good practice to perform a dark calibration every time
you turn on the MK350D, when you move to a different
temperature environment (hot to cold or cold to hot), or
when you move to a different light intensity environment
(light to dark or dark to light).
You also can use PC software or smart device APPs to do dark
calibration.