Service; General Warning Information On Installation Work; Tightening Torques; Troubleshooting - Invacare Stream Service Manual

Hide thumbs Also See for Stream:
Table of Contents

Advertisement

Invacare® Stream

6 Service

6.1 General Warning Information on Installation Work

CAUTION!
Risk of damage to vehicle
Collisions can be caused if the adjusting washers are removed during fitting work to the drive wheels. Adjusting
washers are often fitted between the drive shaft and the wheel hub to even out tolerances. If these adjusting
washers are removed and not replaced again, collisions can be caused.
– Always replace the adjusting washers exactly as they were before you started dismantling.
CAUTION!
Risk of injury and damage to property, if the maximum speed reduction on a mobility device with a lifter does
not function correctly
The mobility device's control unit must reduce the maximum possible speed as soon as the lifter is raised.
– Test the maximum speed reduction for correct function after any maintenance work or modifications
to the mobility device.

6.2 Tightening Torques

CAUTION!
Risk of damage to mobility device due to improperly tightened screws, nuts or plastic connections.
– Always tighten screws, nuts etc. to the stated tightening torque.
– Only tighten screws or nuts which are not listed here fingertight.
The tightening torques stated in the following list are based on the thread diameter for the nuts and bolts for which no
specific values have been determined. All values assume dry and de-greased threads.

6.3 Troubleshooting

6.3.1 Operational Faults

Proceed as follows if you have any problems:
1. First assess the possible cause of the problem using the following table.
2. Check the remote status display. Evaluate the flash error code.
3. Carry out the necessary checks and repairs as recommended in the following table.
The various power modules can be fitted in connection with different remotes in the mobility device. Rectification of
operational faults depends on the power module fitted. The power modules used are described in the corresponding
controls manual.
The tables for rectification of operational faults listed in the following chapters are only an excerpt from the original
manufacturer's manuals. You can obtain the original manuals from Invacare.
14
Thread
M4
M5
M6
M8
M10
M12
M14
M16
Tightening Torque in Nm ±10 %
3 Nm
6 Nm
10 Nm
25 Nm
49 Nm
80 Nm
120 Nm
180 Nm
1608312-B