Problem
... the display "Er :6 :82" appears?
... the display "Er :6 :83" appears?
... the display "Er :6 :84" appears?

10.MAINTENANCE

11.TECHNICAL DATA

11.1 General technical data

Dimensions
• Depth
• Width
• Height
Net weight
Ambient conditions, operation
• Temperature
• Air pressure
• Humidity
Ambient conditions, storage
• Temperature
• Air pressure
• Humidity
An error occurred during the measuring operation.
- Ask the patient to keep still and repeat the measuring operation
• If the error message recurs:
- inform seca Service
An error occurred during the reference measurement.
- Step off the platform
• Interference caused by reflection
- Remove objects from the immediate vicinity of the device (see "Setting up
the device" on page 219).
- Ask people to keep a reasonable distance from the device (see "Setting
up the device" on page 219).
• Interference cause by other ultrasonic emitters
- Increase the distance from other ultrasonic emitters (see "Setting up the
device" on page 219)
• The ambient temperature of the device is too high or too low.
- Set up the device in an ambient temperature between +10 °C and
+40 °C
- Wait for around 15 minutes until the device has adapted to the ambient
temperature
• The temperature sensor is faulty.
- Inform seca Service
The product must be set up carefully and serviced regularly. Depending on
how frequently the product is used, we recommended servicing at intervals of
3 to 5 years.
NOTICE!
Incorrect measurements as a result of poor servicing
Have servicing and repairs carried out exclusively by seca service or
by an authorized service partner.
You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
General technical data
Cause/solution
+10° C to +40° C (50 °F to 104 °F)
30 % to 80 %, no condensation
-10° C to +65° C (14 °F to 149 °F)
0 % to 95 %, no condensation
English
466 mm
434 mm
2270 mm
16.5 kg
700 hPa - 1060 hPa
700 hPa - 1060 hPa
Maintenance • 247