Download Print this page

Philips M1191B Instructions For Use Manual page 9

Reusable spo2 sensors, adult finger glove, pediatric finger glove, neonate hand/foot wrap, adult/pediatric ear clip, infant finger/toe glove
Hide thumbs Also See for M1191B:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Applying Wrap Style Sensors
Neonate Patients (1-4kg)
Apply M1193A Sensor to hand or foot as shown.
Step
1
Position the sensor so that it surrounds the hand or foot with opti-
cal components opposite each other.
2
Holding sensor, stretch strap slightly (not more than 2.5cm (1.0").
3
Insert stretched strap into slot, hold it there while threading end
through latch. If strap is too long, thread it through second latch.
4
Plug sensor into instrument (or into adapter cable if needed).
5
Inspect and change the sensor application site periodically.
Applying Ear-Clip Style Sensors
Adult or Pediatric Patients (>40kg)
Clip M1194A Sensor to ear as shown.
Step
1
To improve perfusion, massage or warm the ear lobe.
2
Clip probe onto fleshy part of the lobe. Do not position probe on
cartilage or pressed against the head. Clip plastic fixing mecha-
nism on top of the ear. The fixing mechanism minimizes artifact
generated by patient motion.
3
Plug sensor into instrument (or into adapter cable if needed).
4
Inspect and change the sensor application site periodically.
Applying the Wrap Style Sensor
Applying the Ear-Clip Style Sensor
7

Advertisement

loading

This manual is also suitable for:

M1191blM1192aM1193aM1194aM1195a