1. SAFETY WARNINGS Simco-Ion recommends that these instructions be read completely before installation or operation is attempted. Failure to do so could result in personal injury and/or damage to the equipment. This instruction manual uses symbols to identify dangerous situations as follows: NOTE –...
2. DESCRIPTION The IQ Easy Sensor Bar is an electrostatic field measuring device packaged with a form-factor similar to a typical static neutralizing bar. This makes the IQ Easy Sensor Bar suitable for use in plastic film and web processing applications. It can monitor processed materials for the presence, polarity and degree of electrostatic charges.
Page 5
Static sensors are typically located at regular intervals along the length of the bar. With this multi-segment static sensor arraignment, the IQ Easy Sensor Bar provides a resolution in static measurement previously unavailable. The sensors themselves use field chopping technology to provide accurate and long term stability of static field measurement.
3. SPECIFICATIONS Power Supply IQ Power Control Station Input Voltage 24 VDC Input Current 0.05A +0.022A/sensor Input Connection 5-pin M12 (A-coded) female receptacle 1 to 2.5 psi (7 to 17 kPa) nominal operating pressure 5 psi (35 kPa) max operating pressure Input Purge Gas 0.14 scfm (4 NL/min) per sensor at 1 psi (5 kPA) 0.22 scfm (6 NL/min) per sensor at 2.5 psi (20 kPa)
4. INSTALLATION Unpacking Carefully remove all equipment from its carton and inspect the contents: • Check that the details on the packing slip correspond to the details of the product received. • Check that the equipment is free from damage. •...
Page 8
Figure 2: IQ Easy Sensor Bar Dimensions IQ Easy Sensor Bar 5201226 Rev. F...
Page 9
NOTE – SENSITIVE ELECTRONICS: Risk of equipment damage. Do not install bar with charged materials present. Bar must already be connected to power supply in presence of charged materials. Avoid sharp blows to bar. Dropping bar may cause damage to sensor element. Bar should be mounted on a structure that is free of vibration.
Electrical Connections NOTE – Turn line voltage power off at IQ Power Control Station before connecting / disconnecting any devices. This is to avoid potential equipment failure and ensure proper digital communication. The IQ Easy Sensor Bar may be connected directly to the Control Station or the Sensor Bar may be connected in conjunction with BPS static neutralizing power supply.
Page 11
Figure 5: IQ Power Connections (Sensor Bar to Control Station through BPS Power Supply) IQ Easy Sensor Bar 5201226 Rev. F...
Page 12
Connecting Sensor Bar to Control Station (standalone or with other devices) Cables used to connect the IQ Easy Sensor Bar to the Control Station have M12 connectors on both ends. Cables are available in a variety of lengths with straight or right angle connectors at the IQ Easy device end (see Parts &...
Page 13
Figure 7: IQ Power Connections (Sensor Bar and IQ Easy Static Bar to Control Station) IQ Easy Sensor Bar 5201226 Rev. F...
Page 14
Figure 8: IQ Power Connections (Sensor Bar and MPS Power Supply to Control Station) IQ Easy Sensor Bar 5201226 Rev. F...
Page 15
Figure 9: IQ Power Connections (Sensor Bar and Fantom to Control Station) IQ Easy Sensor Bar 5201226 Rev. F...
Purge Gas (recommended) NOTE – Purge gas or compressed air used with this device MUST be clean and dry. Dirt, water or oil in the purge gas will damage the sensors. In dirty or dusty environments low pressure purge gas may be used to prevent the sensors from becoming fouled with debris.
Page 17
Paired Bar: The specific static neutralizing bar used with a sensor bar in closed-loop feedback (CLFB) mode. The static neutralizing bar MUST be upstream (before) the sensor bar. Pairing is controlled by connection to “PS COMM 2” on a BPS power supply or by matching the sensor bar address to an IQ Easy static neutralizing bar or Fantom ionizing blower.
Page 18
Data Logging: Available, must be turned on. Firmware Version Main: The firmware revision in the gateway (main) circuit board. Web Voltage Sensor 1,2,3...: The voltage level reported by a given sensor module. Firmware Version Sensor 1,2,3…: The firmware revision in the sensor module circuit board.
5. OPERATION The operation of the sensor bar is controlled through the IQ Power Control Station. In operation, a device icon appears on the IQ Power Control Station, a tab appears on the neutralizer page when the sensor is paired with the neutralizer. Tapping on the device icon / tab for the sensor bar enters a series of pages containing information about that sensor bar.
6. MAINTENANCE NOTE – Turn off power to sensor bar before performing any maintenance. Do not disconnect sensor bar if charged materials are present. Cleaning Clean the sensor bar using a lint-free wiper moistened (but not saturated) with isopropyl alcohol. Use caution around the sensor modules and avoid contact with the sensor aperture.
7. TROUBLESHOOTING Signal Problem Cause Solution Control Station Apply power to Control Station & turn on LEDs on bar No supply is off do not light voltage Wiring problem Replace cable with known good cable Correct problem causing unacceptable Web voltage web voltage Yellow LED Bar is in...
8. PARTS AND ACCESSORIES Item Part Number Mounting Kit 5051777 Cables for connecting Sensor Bar to Control Station with straight connector at bar end 1.5 meter (4.9 foot) 5051939 5 meter (16.4 foot) 5051791 10 meter (32.8 foot) 5051792 20 meter (65.6 foot) 5051793 30 meter (98.4 foot) 5051794...
9. WARRANTY AND SERVICE This product has been carefully tested at the factory and is warranted to be free from any defects in materials or workmanship. Simco-Ion will, under this warranty, repair or replace any equipment that proves, upon our examination, to have become defective within one year from the date of purchase.