Keysight 16194A High Temperature Component Fixture
Operation and Service Manual

3 Operation

Introduction

This chapter describes how to perform error compensation and how to
measure the device with the 16194A. The topics include the following.

Operating Flow

The following shows typical operating procedure. Before performing a
measurement, it is
The error compensation functions are different for each instrument. For
procedure on how to perform each error compensation,
supplied with your analyzer.
1. Perform calibration at the measurement terminal of your analyzer, and
2. Perform electrical length compensation (Fixture Select/Port Extension).
3. Perform SHORT compensation.
4. Perform LOAD compensation, if necessary.
5. Perform OPEN compensation.
6. Measure the device under test.
Operating flow
Connecting the test fixture for use
Placing shorting device, load device, or device under test
Setting the open condition
necessary to compensate for the residual error of the fixture.
connect the 16194A.
For both surface mount device and leaded device measurements, use the
shorting device.
For both surface mount device and leaded device measurements, use the
50 Ω chip resistor as a LOAD device. The reference value for LOAD
compensation is R = 50 Ω/ L = 0 H.
see the Help Manual