• If the compressor is visibly dirty, clean the compressor
before disinfecting it [see: Cleaning the compressor
housing, page 30].
• Moisten a cloth with the disinfectant.
NOTE! Danger of device fault due to liquid penetration.
Never spray any liquids onto the compressor or the power
cord. If liquid does get into the compressor, it must not be
used under any circumstances. Before starting the com-
pressor again, contact the manufacturer or distributor.
• Wipe the outer surfaces of the housing off thoroughly with
the cloth.

4.3 Care of the connection tube

Dry the connection tube after every inhalation session:
• Connect the connection tube to the compressor.
• Switch your compressor on.
• Leave the compressor to work until the air flowing through
the tube has removed all traces of any condensation in the
tube.

4.4 Replacing the air filter

The air filter must be checked at regular intervals (after every
10th application). If it is discoloured brown or grey, or if it is
damp or clogged, it must be replaced.
In all cases, it must be replaced after 200 operating hours (ap-
prox. 1 year).
The air filter cannot be cleaned and then reused!
Take the air filter out:
• Pull the filter holder out of the
compressor. Use a small
screwdriver, for example, to
prise the filter holder carefully
out of the compressor.