Operating menu
Interface linearized
Navigation
Prerequisite
Description
Additional information
Maximum value
Navigation
Prerequisite
Description
User entry
Factory setting
Diameter
Navigation
Prerequisite
Description
User entry
Factory setting
Additional information
154
Setup → Advanced setup → Linearization → Interface linearized
Operating mode (→ 128) = Interface or Interface with capacitance
Displays the linearized interface height.
The unit is defined in the Unit after linearization parameter → 152.
Setup → Advanced setup → Linearization → Maximum value
Linearization type (→ 151) has one of the following values:
• Linear
• Pyramid bottom
• Conical bottom
• Angled bottom
• Horizontal cylinder
• Sphere
Linearized value corresponding to a level of 100%.
–50 000.0 to 50 000.0 %
100.0 %
Setup → Advanced setup → Linearization → Diameter
Linearization type (→ 151) has one of the following values:
• Horizontal cylinder
• Sphere
Diameter of the cylindrical or spherical tank.
0 to 9 999.999 m
2 m
The unit is defined in the Distance unit parameter (→ 128).
Levelflex FMP55 HART
Endress+Hauser
Need help?
Do you have a question about the Levelflex FMP55 and is the answer not in the manual?