3 Commissioning

3.1 Unpacking

The D
P
1000 is supplied packaged in a product-specific carton. The recommended unpacking
OSI
UMP
procedure for the D
OSI
1. Remove the securing straps
2. Open the packaging
3. Remove the inside cover of the packaging
4. Lift the unit out of the packaging. Do not grip the unit by the pump head when lifting it out.
5. Place the unit onto a sturdy table or bench.
6. Remove all miscellaneous packaging items (foam rubber and safety stickers...) and place them
back into the transport packaging since they may be needed if it becomes necessary to transport
the unit again.
WARNING

3.2 Mains connection

The D
P
1000 must be connected to the mains power supply using the enclosed mains cable. The
OSI
UMP
electrical connection ratings must correspond to the ratings specified on the sticker on the rear side of
the unit and may be neither higher nor lower than these values.
Connect the power cable of the D
order to avoid electrical shock. Do not use adapters for the plugs and do not remove the earthing con-
nections from the cable. If it is necessary to use an extension cable, please use only three-core cables
with correctly earthed plugs for this.

3.3 Inserting the tube fitting in the pump

3.4 Metal Bar fixation making the sturdy housing a stable stand

Screw on the metal bar to use the DosiPump housing as a stable stand.
1. Screw the stainless steel stand into the opening on the housing cover.
2. Attach the stainless steel cross-bar with stand cross-clamps.
3. Also firmly clamp the quick fixer clamps.

3.5 Main Switch

Press the master switch on the rear panel of the D
is connected, this printer is also initialised. The software version and type are displayed. You will see the
main menu after initialisation is complete.
Page 8 / 24
P
1000 is as follows:
UMP
If there are signs of condensation on the D
hours for the condensation to dry out before switching on or operating the unit.
P
OSI
UMP
Perform the following steps to insert the tube in the pump:
1. Move the lever on the pump head to the right
2. Insert the tube fitting into the pump
3. Move the lever back to the left
4. Immerse the end of the tube (intake end) into the medium to be dis-
pensed or connect it to the optional culture medium preparation unit.
1000 only to a fully earthed mains power supply system in
P
1000. The unit is then reinitialised. If a printer
OSI
UMP
DosiPump 1000 User Manual
P
1000, allow at least two
OSI
UMP
Version 1.01E