In order to replace the device fuse, use a small screwdriver to
lever the fuse holder of the power receptacle down and out.
Fuse:
3.15 A - 250 V AC - slow-blow
No fuses with other ratings may be used!
WARNING
Faults and errors occurring during operation
Symptom
Fault signal
Pump blocked
Fault signal:
No program entered
Fault signal:
DosiPump barred
No valid license
Page 18 / 24
Always disconnect the mains cable first before replacing the fuse!
Possible cause
The culture medium in the tube is too
viscous. You have used a tube which
is too large.
The selected program has no name
(has not been edited).
License has expired
DosiPump 1000 User Manual
Correction
Remove the tube and run the pump
manually with the "PUMP" menu.
Edit the selected program and assign
it a program name.
Apply for a new license from the sell-
er with the license application code
and enter the license code.
Version 1.01E