DosiPump 1000 User Manual Contents OVERVIEW ............................3 ................. 3 REFACE FURTHER TO THE ANUAL 1.01 ..................3 OSIPUMP ANUAL ERSION 1000 ................4 ASIC INFORMATION ON THE SAFETY INFORMATION ......................... 6 ......................6 PERATING ENVIRONMENT ..................... 6 AFETY CONSCIOUS WORKING .................
DosiPump 1000 User Manual 1 Overview 1.1 Preface further to the User Manual The User Manual explains the mode of operation and the operating procedures for the D 1000. The User Manual is subdivided into five main sections: 1. Overview 2.
DosiPump 1000 User Manual Basic information on the D 1000 1.3.1 Introduction to the D 1000 Peristaltic Pumps Principle Peristaltic pumps work by reduction of a volume within a tube by a series of rollers. The fluid is forced through tubing by a number of rollers, which rotate squeezing a flexible tube against the pump housing + Large Flows range...
Page 5
DosiPump 1000 User Manual • Variably adjustable sample volumes, speed, pause time and AntiDrop function • Automatic initialisation and self-test run • Up to 8 different tube calibrations can be stored and are available in any program • The tube fittings can be exchanged easily thanks to the practical lever closure on the pump head •...
DosiPump 1000 User Manual 2 Safety information 2.1 Operating environment Several factors must be noted when selecting the right operating environment for the unit. The room se- lected may not pose an aggressive environment either for the unit itself or for its intended application. The D 1000 should be operated only under ambient conditions corresponding to a normal envi- ronment (5°C with maximum 80 % relative humidity).
DosiPump 1000 User Manual 2.4 Safety information for the user company The superiors of the operating personnel must have understood and must comply with Chapter "Safety information" and the aspects of operation relevant to safety. They must ensure that the operating per- sonnel have understood and follow the safety information.
DosiPump 1000 User Manual 3 Commissioning 3.1 Unpacking The D 1000 is supplied packaged in a product-specific carton. The recommended unpacking procedure for the D 1000 is as follows: 1. Remove the securing straps 2. Open the packaging 3. Remove the inside cover of the packaging 4.
DosiPump 1000 User Manual 4 Operation of the D 1000 4.1 Fundamentals Direct Dosing; DosiPump offers a direct dispense function straight from the main screen. (4.2) Speed & Time controlled / no absolute volumes (undefined- / not calibrated tube set). This function allows fast, user controlled dispensing of any volume. Programs;...
DosiPump 1000 User Manual 4.2 Direct Dosing MAIN SCREEN PROGRAMS SETTINGS PUMP DIRECTION START / PUMP SPEED Screenshot Step Start Screen Press anywhere to continue DosiPump allows direct dosing straight from the main screen (continuous pump mode). No need for previous program setting. The speed and the pump direction can be selected.
DosiPump 1000 User Manual 4.3 Execute Dosing Program MAIN PROGRAM SCREEN MODE ENTER / EXIT PREVIOUS NEXT PAGE BASKET / DELETE PROGRAM EDIT PROGRAM START PROGRAM Step Screenshot SCROLL FOR Enter the program section by pressing this icon MORE PROGRAMS Start / Stop any selected program PROGRAM START PO-...
DosiPump 1000 User Manual Execution Screen Execution Screen REVOLUTIONS TARGET VOLUME CURRENT SPEED CURRENT VOLUME STOP PROGRAM PAUSE CURRENT DOSE Step Screenshot Before DosiPump starts the selected program the cycle repetition screen appears automatically. Select number or cycles (repetition of complete program) Alternatively select Endless repetitions (repeats until user selected stop) Page 12 / 24...
DosiPump 1000 User Manual 4.5 Edit / Create Program MODE ENTER / EXIT PREVIOUS NEXT PAGE BASKET / DELETE PROGRAM EDIT PROGRAM START PROGRAM Step Screenshot DosiPump allows to type and store up to 20 different SCROLL FOR MORE Programs. PROGRAMS Selects the Program to type/edit TYPE / EDIT...
DosiPump 1000 User Manual 4.6 Calibration of Tubes CALIBRATION SCREEN PROGRAMS SETTINGS PUMP DIRECTION START / PUMP SPEED Step Screenshot Enters into the setting menu Opens the tube calibration menu Select a tube to be calibrated refil the tube and eliminate any air bubbles in the tube.
DosiPump 1000 User Manual 4.7 System Menu System Screen CONTRAST TOUCH TUBE CALIBRATION INTERFACE LANGUAGE INTERFACE ABOUT THIS UNIT DATE / TIME Step Screenshot Tube Calibration See P4.6 4.7.1 Language Select between English / German Menu 4.7.2 Date / Time Setting Set Date / Time Select European / US Display 4.7.3 Touch Screen Calibration...
DosiPump 1000 User Manual 5 Technical information 5.1 Introduction Chapter "Technical information" contains additional information on operation of the D 1000. It is assumed that the unit is operated by a trained D 1000 user with the required access rights. 5.2 Cleaning Cleaning procedures must be performed regularly in order to ensure optimum performance of the 1000.
DosiPump 1000 User Manual 5.4 Calibration Corporate guidelines prescribe that all companies operating with a quality assurance system must regu- larly calibrate their instruments and equipment. Calibration must ensure that the unit complies with the stipulated specifications. When should calibration be carried out? Calibration relates to the dosing volume of the pump as a function of the tube.
Page 18
DosiPump 1000 User Manual In order to replace the device fuse, use a small screwdriver to lever the fuse holder of the power receptacle down and out. Fuse: 3.15 A - 250 V AC - slow-blow No fuses with other ratings may be used! WARNING Always disconnect the mains cable first before replacing the fuse! Faults and errors occurring during operation...
DosiPump 1000 User Manual 5.6 Electrical connections The connections for the mains power supply and the optional interfaces are located on the right-hand side of the D 1000 rear panel. Caption 1. Footswitch Socket DIN41524 5-pin 2. PC / printer connection SUB-D socket 9-pin 3.
DosiPump 1000 User Manual 5.7 Technical data Length 305 mm Depth 275 mm Height 205 mm Net weight 7.1 kg Mains connections 85-132 V AC / 176-264 V AC, 47-63 Hz Mains fuse 3.15 A / 250 V AC slow-blow Dosing range 1.0 –...
DosiPump 1000 User Manual 6 Additional Application Data 6.1 Accessories & Options Accessories and options are subject to alteration For complete and valid selection please demand the domestic price list NEW 2010 DosiPump DP1000, IP64- splash water protected housing made of stainless steel, Touch-Screen &...
DosiPump 1000 User Manual Reproducibility and Precision Data from BioTool test lab (June 25 2010) DP1000 Serial 730.13.00014 DP1000 Fast Serial 740.13.00016 - Test with tab water - No evaporation protection / oil - Balance AND EK-3000i 730000 740000 Page 22 / 24 Version 1.01E...
DosiPump 1000 User Manual 6.3 Certificates Please ask your local supplier for required additional certificates. 6.4 Product Registration Please register your unit with you local supplier in order to get automatic updates and product infor- mation from BioTool AG 6.5 Contacts & Feedback Sales &...
Page 24
DosiPump 1000 User Manual Page 24 / 24 Version 1.01E...