Version:1.0 User Manual Wrist Blood Pressure Monitor TMB-1581-S Wrist Type Thank you very much for selecting a TRANSTEK Wrist Blood Pressure Monitor Guangdong Transtek Medical Electronics Co., Ltd. TMB-1581-S. Zone A, No.105 ,Dongli Road, Torch Development District, To use the monitor correctly and safely, please read the manual thoroughly.
CATALOGUE INTRODUCTION ..................... General Description Indications for Use Contraindications Measurement Principle Safety Information LCD Display Symbol Monitor Components List BEFOREYOU START ..................Installing and Replacing the Batteries Setting Date, Time and Measurement Unit MEASUREMENT ....................Tie the Cuff Start the Measurement DATA MANAGEMENT ..................
The design provides you with two years of reliable such as cardiac pacemakers, defibrillators. service. Readings taken by the TMB-1581-S are equivalent to those obtained by a Measurement Principle trained observer using the cuff and stethoscope auscultation method.
INTRODUCTION INTRODUCTION Safety Information CAUTION The signs below might be in the user manual, labeling or other component. They are the requirement of standard and using. * This device is intended for adult use in homes only. * The device is not suitable for use on neonatal patients, pregnant women,patients with implanted, electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation, peripheral, Symbol for “THE OPERATION Symbol for “TYPE BF APPLIED...
Page 5
INTRODUCTION INTRODUCTION CAUTION CAUTION * The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable * Warning: Do not apply the cuff over a wound;otherwise it can cause further injury. anesthetic mixture with air of with oxygen or nitrous oxide. *Do not inflate the cuff on the same limb which other monitoring ME equipment is applied around * Warning: No servicing/maintenance while the ME equipment is in use.
* If you have any problems with this device, such as setting up, maintaining or using, please contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself in the event of malfunctions. The device must only be serviced, repaired and opened by individuals at authorized sales/service centers.
INTRODUCTION INTRODUCTION Monitor Components SYMBOL DESCRIPTION EXPLANATION Measurement unit of the blood pressure (1kPa=7.5mmHg) Measurement unit the blood pressure mmHg (1mmHg=0.133kPa) Blood pressure monitor is detecting an Irregular heartbeat irregular heartbeat during measurement. SYSTOLIC Current time Month/Day/Year,Hour : Minute LCD DISPLAY Blood pressure Indicate the blood pressure level level indicator...
BEFORE YOU START BEFORE YOU START Setting Date, Time and Measurement Unit It is important to set the clock before using your blood pressure Replace the batteries whenever the below happens monitor, so that a time stamp can be assigned to each record that is stored in the memory.
Page 10
BEFORE YOU START BEFORE YOU START .Repeat step 2 and 3 to confirm [HOUR] and [MINUTE]. 3. When you get the right year, press “START/STOP” button to confirm your selection and it will turn to the next step. .Repeat step 2 and 3 to confirm the [MEASUREMENT UNIT]. 4.Repeat step 2 and 3 to confirm [MONTH] and [DAY].
MEASUREMENT MEASUREMENT Start the Measurement Tie the Cuff 1.When the monitor is off, press “START/STOP” button to turn on the monitor, Remove all accessories (watch, bracelet,etc) from your wrist. If your physician has diagnosed you with and it will finish the whole measurement. poor circulation in your wrist, use the other one.
DATA MANAGEMENT DATA MANAGEMENT MEASUREMENT Recall the Records When the monitor is off, press “MEM” button to show the average value of the latest three measurement records.If the records are less than 3 groups,it will display the latest record It means the total The corresponding The corresponding instead.
DATA MANAGEMENT DATA MANAGEMENT Delete the Records If you did not get the correct measurement, you can delete all 3.If there is no record, the results by following steps below. right display will be shown. Hold pressing “MEM” button for 3 seconds when the monitor is in the memory recall mode, the “dEL All”...
INFORMATION FOR USER INFORMATION FOR USER Maintenance Tips for Measurement To obtain the best performance, please follow the instructions below. Measurements may be inaccurate if taken in the following circumstances. Within 1 hour Immediate measurement Avoid shaking and collision. Put in a dry place and Avoid immersing it in the water.
ABOUT BLOOD PRESSURE ABOUT BLOOD PRESSURE Irregular Heartbeat Detector What are systolic pressure and diastolic pressure? Systolic Diastolic When ventricles contract and pump blood An irregular heartbeat is detected when a heartbeat rhythm varies while the device is out of the heart, blood pressure reaches its blood entering blood discharging measuring systolic pressure and diastolic pressure.
ABOUT BLOOD PRESSURE TROUBLESHOOTING Why does my blood pressure fluctuate throughout the day? This section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor. If the product is not 1.
SPECIFICATIONS TROUBLESHOOTING PROBLEM SYMPTOM CHECK THIS REMEDY Power supply Battery powered mode: 2*AAA batteries Display mode Digital LCD display V.A.32mmx45mm Retake the measurement.If the problem Error A calibration error EExx,shows on persists,contact the retailer or our customer Measurement mode Oscillographic testing mode message occurred.
Contact Information Mode of operation Continuous operation Degree of protection For more information about our products, please visit www.transtek.cn.you can get customer Type BF applied part service, usual problems and customer download, transtek will serve you anytime. Device Classification Internally Powered ME Equipment IP22: The first number 2: Protected against solid foreign objects of 12,5mm Ф...
COMPLIED STANDARDS LIST COMPLIED STANDARDS LIST Complied Standards List EN 1060-3:1997+A2:2009 Non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood EN ISO 14971:2012 / ISO 14971:2007 Medical devices - Risk management pressure measuring systems Application of risk management to medical devices IEC 80601-2-30:2009+A1:2013 Medical electrical equipment- Part 2-30: Particular requirements for the basic safety and EN ISO 15223-1:2016 / ISO 15223-1:2016 Medical devices.
Warning:Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of Harmonic emissions the equipment TMB-1581-S, including cables specified by the manufacturer. Otherwise, Not application IEC 61000-3-2 degradation of the performance of this equipment could result.
Page 21
EMC GUIDANCE EMC GUIDANCE Table 2 Guidance and manufacturer’s declaration – electromagnetic Immunity Power frequency 30 A/m 30 A/m magnetic field 50Hz/60Hz 50Hz/60Hz IEC 61000-4-8 Immunity Test IEC 60601-1-2 Compliance level Conduced RF Not application Test level Not application IEC61000-4-6 Electrostatic ±8 kV contact ±8 kV contact...
Page 22
EMC GUIDANCE EMC GUIDANCE Table 3 GSM 1800; Pulse Guidance and manufacturer’s declaration - electromagnetic Immunity 1720 1700- CDMA 1900; modulation b) 1990 GSM 1900; 217Hz IMMUNITY Radiated RF Test Band Service Modulation Modulation Distance (m) 1845 DECT; TEST IEC61000-4-3 Frequency (MHz) LTE Band 1,...