Page 1
Blood Pressure Monitor TMB-1591-BS Arm Type FCC ID:OU9TMB1591BS Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB-1591-BS. Guangdong Transtek Medical Electronics Co., Ltd. To use the monitor correctly and safely, please read the manual Zone B, No.105 ,Dongli Road, Torch Development District, Zhongshan,528437,Guangdong,China thoroughly.
CATALOGUE CATALOGUE Table of Contents INTRODUCTION..................2 General Description Safety Information Indications for Use Contraindications Measurement Principle LCD Display Signal Monitor Components BEFORE YOU START................8 The Choice of Power Supply Installing and Replacing the Batteries MEASUREMENT..................9 Pairing Tie the Cuff Start the Measurement INFORMATION FOR USER..............14 Tips for measurement Maintenances...
The design provides you with two years of reliable service. Symbol for “ENVIRONMENT Caution: These notes must be Readings taken by the TMB-1591-BS are equivalent to those PROTECTION – Electrical waste observed to prevent any damage products should not be disposed obtained by a trained observer using the cuff and stethoscope to the device.
Page 4
SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself in the event of * When not in use, store the device with the adapter in a dry room and protect it against extreme moisture, malfunctions.
Battery Indicator Indicate the current battery List Grade The grade of the blood pressure 1. Blood Pressure Monitor 2. Cuff (about 22cm~42cm) (TMB-1591-BS) (Type BF applied part) Shocking reminder Shocking will result in inaccurate Data transmission error Data transmission error Heartbeat Heartbeat dectetion during measurement 3.
BEFORE YOU START MEASUREMENT The Choice of Power Supply Pairing 1.Battery powered mode: Turn on Bluetooth and APP. Make sure both are ON when pairing is 6VDC 4*AA alkaline batteries proceeding. 2.AC adaptor powered mode: When the blood pressure monitor is off, press On/Standby button for (Can be supplied by AC adaptor model 2 full seconds to enter bluetooth pairing mode.
MEASUREMENT MEASUREMENT Tie the Cuff Start Measurement When the blood pressure monitor is off, press On/Standby button to 1.Tie the cuff on your upper arm, turn it on, it will finish the whole measurement. then position the tube off-center toward the inner side of arm in line with the little finger.
Page 8
MEASUREMENT MEASUREMENT The blood pressure monitor will proceed to data transmission after During the measurement, if you press the On/Standby button to stop measurement. The bluetooth symbol flashes on the LCD indicates the measurement, the numerics are cleared. It will display as below: data is transmitting.
INFORMATION FOR USER INFORMATION FOR USER Tips for Measurement Maintenance Measurements may be inaccurate if taken in the In order to get the best performance, please follow the following circumstances. instructions below. Immediate measurement wait at least 1 hour after dinner after tea, coffee, smoking Avoid touching water, Put in a dry place and avoid the...
ABOUT BLOOD PRESSURE ABOUT BLOOD PRESSURE What are systolic pressure and diastolic Why does my blood pressure? Systolic Diastolic pressure fluctuate When ventricles contract and pump blood out of the heart, the blood pressure blood discharging blood entering throughout the day? vein artery reaches its maximum value in the cycle,...
TROUBLESHOOTING SPECIFICATIONS This section includes a list of error messages and Battery powered mode: frequently asked questions for problems you may 6VDC 4*AA alkaline batteries encounter with your blood pressure monitor. If the Power supply AC adaptor powered mode:6V products not operating as you think it should, check here (Can be supplied by AC adaptor model before arranging for servicing.
This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. For more information about our products, please visit www.transtek.cn.you can get customer service, usual problems and customer download, transtek will serve you anytime. Manufactured by: Guangdong Transtek Medical Electronics Co., Ltd.
COMPLIED STANDARDS LIST EMC GUIDANCE Complied Standards List EMC Guidance EN ISO 14971:2012 / ISO 14971:2007 Medical devices - 1)This product needs special precautions regarding EMC and needs to be Risk management Application of risk management to medical devices installed and put into service according to the EMC information provided, EN ISO 15223-1:2016 / ISO 15223-1:2016 Medical devices.
EMC GUIDANCE EMC GUIDANCE Table 2 Table 3 Guidance and manufacturer’s declaration – electromagnetic immunity Guidance and manufacturer’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. The device is intended for use in the electromagnetic environment specified below.
Page 15
EMC GUIDANCE EMC GUIDANCE Table 5 Table 4 Guidance and manufacturer’s declaration - electromagnetic immunity Recommended separation distances between portable and mobile RF communications equipment and the device. The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device, should assure that it is used in such an environment.
Need help?
Do you have a question about the TMB-1591-BS and is the answer not in the manual?
Questions and answers