SoilViewer Warning Messages
There are warning signals programmed into the SoilViewer to warn the operator that data are not being
recorded, so that corrective action can be taken. If data aren't being recorded, a warning beep will sound
from the computer, and the text indicator of the data that is missing information will blink. Fr example, if the
DGPS isn't being received (or the NMEA string containing speed) the Fix indicator text will blink. If EC
values are negative, they will also blink. Additional messages may appear in the Warning Message Box as
follows:
Figure 4
Potential SoilViewer Warning Messages
Error EC_RPBT - Unit Raised out of ground, but EC readings are positive. Check for a moist soil buildup or
a pinched cable causing a short between EC electrodes.
Error EC_NP - Row unit is engaged in soil, but EC readings are negative. Check for continuity, and make
sure EC electrodes are penetrating soil at least 2".
Error EC_N30 - Excessive Negatives in EC data, check coulter continuity and soil contact.
Error OM_DPT - Row unit not penetrating soil enough. Depth should be 1.5" or > when mapping.
Error EC_EXMD - Excessive noise on EC readings; check continuity.
Error OM_LW - OM readings are too low. Check wearplate window for clarity.
Error_No Comm.- Check USB/serial or Bluetooth connections to V-Sense Controller. Unplug USB or re-pair
bluetooth connectivity and restart PC if communication does not reestablish.
Error No_GPS - Check GPS settings (4800 baud, 1 hz, and only VTG and GGA Strings enabled)