Quick Start Guide
2. To keep the transmitter enclosure watertight, install plugs (provided)
in the unused conduit openings.
3. Install the system in an area where vibrations and electromagnetic
and radio frequency interference are minimized or absent.
4. Be sure there is easy access to the transmitter and sensor(s).
2.3
Sample requirements
Be sure the sample meets the following requirements:
1. Temperature: 32 to 122 °F (0 to 50 °C )
2. Pressure: 3 to 65 psig (122 to 549 kPa abs)
3. Minimum flow: 3 gal/hr (11 L/hr)
2.4
Mounting, inlet, and drain connections
The Rosemount
Refer to
cell adapters as shown in the figures. For Rosemount FCL-02 (free chlorine
with continuous pH adjustment), you must also install the pH sensor.
8
CAUTION
Hazardous areas
The system is not suitable for use in hazardous areas.
™
FCL is intended for wall mounting only.
Figure 2-1
or
Figure 2-2
for details. The sensor(s) screw into the flow