Figure 3-1: Analog output connections
The analog outputs are on the main board near the hinged end of the
enclosure door.
For best EMI/RFI protection, use shielded output signal cable enclosed in
earth-grounded metal conduit.
Keep output signal wiring separate from power wiring. Do not run signal and
power or relay wiring in the same conduit or close together in a cable tray.
3.3
Alarm wiring
The alarm relay terminal strip is located just below the power connector on
the power supply board.
See
Figure
Quick Start Guide
3-2.