Keep alarm relay wiring separate from signal wiring. Do not run signal and
power or relay wiring in the same conduit or close together in a cable tray.
3.4
Wire sensor
The Rosemount
transmitter. If it is necessary to replace the sensor cable, refer to the
instructions below.
Procedure
1. Shut off power to the transmitter.
2. Loosen the four screws holding the front panel in place and let it drop
down.
3. Locate the appropriate signal board.
Slot 1 (left)
communication
4. Loosen the gland fitting and carefully push the sensor cable up
through the fitting as you pull the board forward to gain access to the
wires and terminal screws.
5. Wire the sensor to the signal board.
Refer to the wiring diagrams in
Quick Start Guide
FCL is provided with sensor cables pre-wired to the
Slot 2 (center)
input 1 (chlorine)
Slot 3 (right)
input 2 (optional)
Figure 3-3
and
Figure
Quick Start Guide
3-4.
15