Emerson Rosemount FCL Series Quick Start Manual page 5

Free chlorine system with rosemount 1056 transmitter
Hide thumbs Also See for Rosemount FCL Series:

Advertisement

June 2019
1
Specifications
Rosemount 1056 Transmitter
For Rosemount 1056 Transmitter specifications, see the Rosemount 1056
Transmitter Reference Manual on Emerson.com/Rosemount: Manual:
Rosemount 1056 Dual-Input Transmitter.
Table 1-1: General Specifications
Characteristic
Sample requirements
Sample conductivity
Process connection
Drain connection
Wetted parts
Response time to step change in chlorine
concentration
Quick Start Guide
Specification
Pressure: 3 to 65 psig (122 to 549
kPa abs). A check valve in the inlet
prevents the sensor flow cells from
going dry if sample flow is lost. The
check valve opens at 3 psig (122 kPa
abs). If the check valve is removed,
minimum pressure is 1 psig (108 kPa
abs).
Temperature: 32 to 122 °F (0 to 50
°C)
Minimum flow: 3 gal/hr (11 L/hr)
Maximum flow: 80 gal/hr (303 L/hr);
high flow causes the overflow tube
to back up.
>50 µS/cm at 77 °F (25 °C)
¼-in. OD tubing compression fitting (can
be removed and replaced with barbed
fitting for soft tubing)
¾-in. barbed fitting. Sample must drain
to open atmosphere.
Overflow sampler and flow cell: acrylic,
polycarbonate, Kynar
silicone
Chlorine sensor: Noryl
silicone, polyethersulfone, polyester, and
platinum
pH sensor (Rosemount
Stainless steel, glass, Teflon
polyphenylene sulfide, EPDM, and
silicone
< 80 sec to 95% of final reading for inlet
sample flow of 3 gph (11 L/hr)
Quick Start Guide
®
, nylon, and
®
®
, Viton
, wood,
3900VP):
®
,
5