ResMed H5i Information Manual page 4

Positive airway pressure devices, heated humidifier, air tubing.
Hide thumbs Also See for H5i:
Table of Contents

Advertisement

Available languages

Available languages

Problem/possible cause
SmartStart/Stop is disabled.
Note: SmartStart/Stop is not available in
S9 Escape and S9 Escape Auto.
Device does not stop when you remove your mask
SmartStart/Stop is disabled.
Note: SmartStart/Stop is not available in
S9 Escape and S9 Escape Auto.
SmartStart/Stop is enabled but the device does not stop automatically when you remove your
mask
Incompatible mask system being used.
The patient is using a nasal pillows
mask with a set pressure less than
7 cm H
O.
2
Pressure rises inappropriately
Talking, coughing or breathing in an
unusual manner.
Mask cushion is buzzing against the
skin.
Displays error message: Heated tube fault, replace tube
Device has been left in a hot
environment.
There is a fault in your ClimateLine or
MAX
ClimateLine
Displays error message: Humidifier fault, replace humidifier
Device has been left in a hot
environment.
There is a fault in your H5i heated
humidifier.
4
air tubing.
Solution
Enable SmartStart/Stop.
Enable SmartStart/Stop.
Only use equipment recommended by ResMed.
Disable SmartStart/Stop.
Avoid talking with a nasal mask on, and breathe as normally
as possible.
Adjust the headgear.
Allow to cool before re-use. Disconnect the power cord
and then reconnect it to restart the device.
Discontinue using your ClimateLine or ClimateLine
tubing and contact your clinician/service provider. Use
Standard or SlimLine tubing in the interim.
Allow to cool before re-use. Disconnect the power cord
and then reconnect it to restart the device.
Discontinue using your H5i and contact your clinician/
service provider.