Eclipse 5
Pulse Dose Mode and autoSat Features
Your Eclipse 5 has a feature, called autoSAT
delivers a consistent-sized bolus of oxygen up to
40 breaths per minute (see Max Breath Rate table
below). As your breath rate increases, ie. during
ambulation, the autoSAT feature servo-controls
the device, delivering the set bolus size. autoSAT
Technology is proprietary to CAIRE and is utilized
during Pulse Mode only.
Max Breath Rate
Bolus
Size
Pulse
Dose
Setting
15%)
mL
1.0
16
2.0
32
3.0
48
4.0
64
5.0
80
6.0
96
7
128
8
160
9
192
Note: Bolus volume decreases as breath rate
exceeds published range.
Note: The pulse dose setting number (1-9) is not
equal to LPM.
WARNING: PULSE DOSE MODE
SETTINGS MUST BE DETERMINED
FOR EACH PATIENT INDIVIDUALLY
FOR THEIR NEEDS AT REST, DURING
EXERCISE AND WHEN TRAVELING.
WARNING: THE ECLIPSE 5 MAY NOT BE ABLE TO
DETECT ALL INSPIRATORY EFFORTS IN PULSE
MODE. IF THE ECLIPSE 5 DOES NOT RESPOND TO
YOUR INSPIRATION EFFORTS, ASK YOUR HOME
CARE PROVIDER TO CHECK THE SENSITIVITY
SETTING.
18 - ENG
PN 21182453-C4 E | User Manual
that
®
AC Power
Supply &
DC Power
Power
Supply
Cartridge
(Battery)
Max Breath
Max Breath
Rate
Rate
40
40
40
40
40
40
40
31
37
25
31
20
23
15
18
12
15
10
Note: To Equipment Provider: The following oxygen
administration accessories are recommended for
use with the Eclipse 5:
• Nasal Cannula: CAIRE Part Number 5408-SEQ
• Firebreak: CAIRE Part Number 20629671
A firebreak is required for use with any cannula.
• CAIRE offers a firebreak intended to be used
in conjunction with the oxygen concentrator. The
firebreak is a thermal fuse to stop the flow of gas in
the event that the downstream cannula or oxygen
tubing is ignited and burns to the firebreak. It is
placed in-line with the nasal cannula or oxygen
tubing between the patient and the oxygen outlet
of the Eclipse 5. For proper use of the firebreak,
always refer to the manufacturer's instructions
(included with each firebreak kit).
• For any additional recommended accessories,
please see the Accessories Catalog (PN ML-
LOX0010) available on www.caireinc.com.
Step 5:
Adjust the Flow Setting to
the Prescribed Level
Using the Flow Setting Buttons, adjust the flow
setting to the prescribed setting supplied by your
physician.
WARNING: IT IS VERY IMPORTANT
TO SELECT ONLY THE PRESCRIBED
LEVEL OF OXYGEN. DO NOT CHANGE
THE FLOW SELECTION UNLESS YOU
HAVE BEEN DIRECTED TO DO SO BY A
LICENSED CLINICIAN.
Step 6: Begin Using Your Eclipse 5
Breathe normally through the nasal cannula.
Step 7: Universal Cart Operation
The Universal Cart is designed for use on pedestri-
an-type surfaces.
Align the locater holes on the bottom of the device
to the tabs on the Universal Cart.
Align the large threaded screw to the insert at
the back of the Eclipse 5. Then hand-tighten the
knob on the cart until the Universal Cart is tightly
secured to the device.
Depress the push-button on the
Universal Cart handle to adjust
the height of the handle.