Objective
To demonstrate the capability of the V3500A handheld power meter for average pulse power measurement.
Instructions
On the ESG :
1. Set the instrument to its default
setting.
2. Set the frequency and
amplitude.
3. Set the pulse signal.
4. Activate the pulse signal.
5. Turn ON the RF Output
Instructions
On the V3500A :
1. Turn on the V3500A.
2. Set the frequency to 1000 MHz. Press [
3. Set the average count to 4.
4. Read the measurement.
Keystrokes
Press [Preset].
Press [Frequency] 1000 MHz.
Press [Amplitude] 0 dBm.
Press [Pulse], {Pulse Period} 40 µs and {Pulse Width} 20 µs.
Toggle {Pulse On/Off} to "ON".
Toggle [RF On/Off] to "ON".
Keystrokes
Press [
] ON/OFF button to turn on the V3500A.
] OR [
Press [
] button to enter menu mode.
Press [
] or [
Press [
] or [
Press [
] button to exit.
Press [
] to exit menu. The measurement, –3.04 dBm displayed at left
hand corner of the display.
7
] to change frequency to 1000 MHz.
] until "Averaging mode" menu is displayed.
] to increase averaging count to <4>.