Failure to observe the following precautions could result in serious personal injury: WARNING Electrical hazard When service or maintenance is performed on an open instrument, there is a risk of electrical shock. Failure to observe the following precautions could result in damage to the equipment: Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 5 / 76...
Page 6
On Customers site, during exchange or maintenance of electronic card, the service personnel must wear an ESD coat and use disposable wristband. NOTICE Check that all relevant legislation and safety standards are complied with before the Sensistor Sentrac is taken into service. 6 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Read this Manual carefully before putting your instrument into service. When reading, pay particular attention to the WARNINGS, CAUTIONS and NOTICES found throughout the text. 2.1 About this Document This document is only intended for INFICON service personnel working with the Sensistor Sentrac Leak Detector. 2.2 Document History Revision...
3 | Components INFICON 3.4 Interior View (Desktop Model) Description Slot for Expansion Module Slot for SD Card USB Port Communication Port Device Port Speaker Front Panel Instrument Casing LED Lamp Power Supply 18 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Page 20
3 | Components INFICON Description Slot for Expansion Module Slot for SD Card USB Port Device Port Connection Port Speaker Front Panel Instrument Casing LED Lamp Power Supply Display Accessories Plate Main PCB 20 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Sensor signal Broken sensor or Change sensor or error, voltage window present baseline voltage broken probe. probe. too low. on screen. Blue level too low, Change sensor LED flashing. below 300 mV. or probe. 22 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Page 23
Try to on screen, Blue between probe 2) and/or Probe restart the LED flashing. and instrument. CANL (Pin 3) wire instrument or broken. change the probe cable. If the problem remains contact service. Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 23 / 76...
Page 24
PCB. service on screen, Blue recommended. LED flashing. Probe hardware Red measure Probe PCB Probe PCB Change Probe error, service window present broken. broken. PCB. recommended. on screen, Blue LED flashing. 24 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
On Customer site, during exchange or maintenance of electronic card, the service personnel must wear an ESD coat and use disposable wristband. CAUTION Do not tighten screws with too much torque. This can damage threads on aluminum parts. Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 25 / 76...
Spanner for notched nuts Part no. 598-438 (LEMO Part no. DCH.91.161.PA) Torx TX10 Torx TX20 Sensor Key Part no. 598-147 Heat Gun Thread locking and adhesive LOCTITE® 2700 or 2701, 2400, 406 and 222 Scissor 26 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Page 27
Calibration Leak/Gas 10 ppm H2 in N2 or calibration leak around 5x10-4 mbarl/s tracer gas (5% H2 in N2) PC (Windows) USB flash drive. Sensistor Sentrac Software Part no. 591-964 Installer USB flash drive. P60 software Installer Part no. 591-964 Display glass adhesive alignment tool Part no.
Check the resistance between the ground pin in the mains socket and the outside of the probe cable connector, should be 9 ohm or lower. If not, check the lock washer on the Probe Connection Port. 28 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Remove the five screws (A) holding the Accessories Plate (B). Remove the Accessories Plate (B) Assemble Place the Accessories Plate (B). Reinstall and tighten the five screws (A). Reinstall the cover, see Assemble in Remove the Cover [} 28]. Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 29 / 76...
6.3x32 mm Disassemble Disconnect the mains connector. Remove the lid (A) containing the fuses. Remove the fuses. Assemble Install the new fuses. The upper fuse is spare. Reinstall the lid (A) containing the fuses. 30 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Install the new Power Supply. Connect the power supply cabling according to the drawing below. Reinstall and tighten the two screws, 4 mm, (A) holding the Power Supply. Reinstall the cover, see Assemble in Remove the Cover [} 28]. Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 31 / 76...
Remove the Battery cabling from the power switch panel and main PCB. Assemble Install the new Battery. Mount the Battery cabling according to drawing below. Reinstall and tighten the four screws (A) holding the Battery. Reinstall the cover, see Assemble inRemove the Cover [} 28]. 34 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
5.7 Changing Main PCB Part Part no. Consumables Main PCB Sensistor 591-823 Sentrac CAUTION Be very cautions when removing the locking on connector J2 (FFC cable, LCD data, on main PCB). See Drawings [} 71]. 36 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Page 37
ð SD-card reader is working. If any of the above tests fails, see Troubleshooting [} 21]. Reinstall the Accessories Plate (B), see Assemble in Remove Accessories Plate [} 29]. Reinstall the cover, see Assemble in Remove the Cover [} 28]. Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 37 / 76...
When function test is performed on an open instrument, there is a risk of electrical shock. CAUTION Be very cautions when removing the locking on connector J2 (FFC cable, LCD data, on main PCB). See Drawings [} 71]. 38 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Page 39
Connect all cables between the new Front Panel and the PCB. Pay special attention to J2. Do not use force! Perform a function test before the assembly continues. Ensure the following functions, check that the: Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 39 / 76...
Page 40
(the shield) and one of the screws in the back. The resistance should be 9 ohm or lower. If any of the above tests fails, see Troubleshooting [} 21]. Reinstall the cover, see Assemble in Remove the Cover [} 28]. 40 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Be very cautious when removing the locking on connector J2 (FFC cable, LCD data, on main PCB). See Drawings [} 71]. Disassemble Disconnect the mains connector. Remove the cover, see Disassemble in Remove the Cover [} 28]. Remove the four screws (A) holding the Front Panel (B). Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 41 / 76...
Page 42
Perform a function test before the assembly continues. Ensure the following functions, check that the: ð Display is centered in the Front Panel. If not, loosen the screws holding the display and align it. 42 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Page 43
(the shield) and one of the screws in the back. The resistance should be 9 ohm or lower. If any of the above tests fails, see Troubleshooting [} 21]. Reinstall the cover, see Assemble in Remove the Cover [} 28]. Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 43 / 76...
When function test is performed on an open instrument, there is a risk of electrical shock. CAUTION Be very cautious when removing the locking on connector J2 (FFC cable, LCD data, on main PCB). See Drawings [} 71]. 44 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Page 45
(the shield) and one of the screws in the back. The resistance should be 9 ohm or lower. If any of the above tests fails, see Troubleshooting [} 21]. Reinstall the cover, see Assemble in Remove the Cover [} 28]. Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 45 / 76...
J2. Carefully tap the screwdriver / locking alternately left/right to loosen the cable from the connector. Do not use force! Remove the Front Panel. Remove the four screws (C) holding the Speaker (D). 46 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Page 47
(the shield) and one of the screws in the back. The resistance should be 9 ohm or lower. If any of the above tests fails, see Troubleshooting [} 21]. Reinstall the cover, see Assemble in Remove the Cover [} 28]. Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 47 / 76...
2400 Disassemble Remove the screw (A) holding the Rubber Foot (B). Remove the Rubber Foot (B). Assemble Install the new Rubber Foot. Reinstall the screw (A) holding the Rubber Foot. Use thread locking compound. 48 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
LOCTITE 2400 Lower Part Disassemble Disconnect the Probe from the Sensistor Sentrac. Use a Spanner for notched nut and slightly loosen the C21 connection port nut (A). Remove the two screws and the lock washers (B). Carefully remove the lower part of the Probe handle (C).
Page 50
Measure the resistance between the Probe tip and the C21 connection port. Resistance should be lower than 9 ohm. Re-tension the C21 connection port nut if necessary. Connect the Probe to the Sensistor Sentrac. Start the Probe and make sure that it works properly, check: ð Sensitivity ð...
LOCTITE 2400 display Disassemble Disconnect the Probe from the Sensistor Sentrac. Open the Probe Cover, see Disassemble in Open Probe P60 Cover [} 49]. Remove all cables on the Probe PCB (A). Remove the screw (B) holding the Probe PCB (A).
From dec 2017 glass P60 Disassemble Disconnect the Probe from the Sensistor Sentrac. Open the Probe Cover, see Disassemble in Open Probe P60 Cover [} 49]. Remove all cables on the Probe PCB (A). Remove the screw (B) holding the Probe PCB (A).
Page 53
Firmly apply the OLED-display seal (E) on the Probe Display Glass (C). Use a scalpel and gently remove the center piece in the OLED-display seal. Reinstall the Probe PCB (A). See Assemble in Changing Probe PCB with Display [} 51]. Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 53 / 76...
591-801 LOCTITE Disassemble Disconnect the Probe from the Sensistor Sentrac. Open the Probe Cover, see Disassemble in Open Probe P60 Cover [} 49]. Remove the LED Lamps connector from the PCB (B). Remove the LED Lamps (A). Carefully tap them out of their sockets.
Part Part no. Consumables Button 598-181 Disconnect the Probe from the Sensistor Sentrac. Disassemble Open the Probe Cover, see Disassemble in Open Probe P60 Cover [} 49]. Remove the screw (B) holding the Probe PCB (A). Loosen the Probe PCB (A).
LOCTITE 2700 or 2701 Disassemble Disconnect the Probe from the Sensistor Sentrac. Open the Probe Cover, see Disassemble in Open Probe P60 Cover [} 49]. Remove the Probe Sensor (A) from the Probe Neck (B) using a Sensor Key. Remove the Probe PCB (D).
O-ring 8x1 NBR70 591-284 Disassemble Disconnect the Probe from the Sensistor Sentrac. Use a Sensor Key to remove the Probe Sensor Nut (A) from the Probe Neck (B). Use a small slotted screw driver to remove the O-ring inside the Probe Sensor Nut (A).
Consumables O-ring 8x1 NBR70 591-284 Disassemble Disconnect the Probe from the Sensistor Sentrac. Use a Sensor Key to remove the Probe Sensor Nut (A) from the Probe Neck (B). Remove the Probe Neck O-ring (C) from the Probe Neck (B).
Part no. Consumables Sensor 590-292 Disassemble Disconnect the Probe from the Sensistor Sentrac. Use a Sensor Key to remove the Probe Sensor Nut (A) from the Probe Neck (B). Pull out the Sensor (C) from the Probe Neck (B). Assemble Insert the new Probe Sensor (C) into the Probe Neck (B).
Turn the Sensistor Sentrac unit on. Verify that the Hand Probe P60 has the same software version number as stated in the Sensistor Sentrac Software Installer. If not: update the probe via USB cable. Desktop Model: connect the power cable and verify that the plug is properly mounted.
Page 61
ð Connect the headphones ð Listen to verify that audio quality sounds ok. ð Turn down the volume on the Sensistor Sentrac unit to verify that the headphone volume decreases as well. Expose the hand probe to 10 ppm hydrogen. Verify that the bar is increasing.
If model is battery operated: ð Charge battery. ð Verify that battery is fully charged. 5.21.2 Visual verification of Sensistor Sentrac Battery surface finish ► Verify that the surface finish looks good. Pay extra attention to the paint finish around the headphone outlet.
Connect the probe to the instrument if the probe-software is to be updated. Double click on the icon on the desktop. Check that the Sensistor Sentrac serial number appear in the COM port window. If the serial number is present, the communication between the computer and the Sentrac is set up.
• Time counter in ms (milli seconds) • A/D value (sensor signal) raw (24 bit) or in mV (depending on settings) • Locating Value • Measuring Value in mV/s (uncalibrated and unconverted) • Data Checksum Output frequency 1-50 Hz. 64 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Conversion value is set to 1 when ppm is selected as calibration unit and/or measuring Unit. Measure Unit less Measuring Mode or Measuring result value Diagnosis / Service Screen / Signals / Measuring value Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 65 / 76...
Page 66
Use full when adjustment have to be done due to environmental circumstances, e.g. accumulation measuring. Displayed measuring unit / Displayed measuring unit value / Calibration Unit Calibrating unit value cc/s Cc/min SCCM 31536000 oz/yr 1112400 66 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
Page 67
Measuring displayed unit: Mass flow rate (e.g. g/yr. oz/yr) is selected as Calibration unit and Measuring displayed unit Mass flow rate is selected as Calibration Unit and when normalized flow rate as Measuring displayed unit: Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 67 / 76...
Page 68
Output signal from processed signal signal Locating value Diagnosis / Service Screen / Signals / Locating result Locating value (numeric unit less) Locating graph Diagnosis / Service Screen / Locating Locating result Graph (Graphic diagram) 68 / 76 Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
See the diagram 1. 6.5 Settings Reset CAUTION All settings affecting measurements will be reset to factory defaults. Click Diagnosis >> Reset >> Settings Click the green Reset button. Wait for the Sensistor Sentrac to restart. Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901) 69 / 76...
Click Diagnosis >> Reset >> Factory Default. Click the green Reset button. Fill in the service password. Click the green Reset button. Restart the Sensistor Sentrac. 6.7 Reset Calibration To reset an erroneous calibration: Click Diagnosis >> Reset >> Calibration.