Electrical Specifications

Specifications

are
temperature of 23°C ± 3°, relative humidity of
40 to 60% and a supply voltage of 6 V ± 0.1 V.
Ranges:
2000 µΩ
Display:
1999
Resolution:
1 µΩ
Test Current:
10 A
Ranges:
2000 mΩ
1999
Display:
Resolution:
1 mΩ
Test Current:
100 mA
Accuracy:
2000µ
range:
0.25% of reading typical, 0.5% max ± 2 digits.
20–200–2000 m
, 20–200
0.25% of reading typical, 0.5% max ± 1 digit.
Power Supply Specifications
Power Supply: Rechargeable 6 V, 7 Ah NiCad
battery; built-in 110/220 V (47-450 Hz) charger;
automatic shutoff after 4 minutes non-use.
Battery Life (Typical):
2000µ
range:
200 to 300 8-second measurements
20-200 m
ranges:
900 30-second measurements (approx.)
2000 m
, 20-200
ranges:
5000 30-second measurements (approx.)
Operating Voltage: 4.5–7 V DC
Recharge Time:
14 hours from full discharge (approx.)
Mechanical Specifications
Connections: Terminals accept 4 mm banana
plugs or 6 mm (min) spade lugs
Display: 2000 count, 7 segment LCD
10/27/99
REPAIR & CALIBRATION MANUAL
given
for
an
ambient
20 mΩ
200 mΩ
19.99
199.9
10 µΩ
100 µΩ
1 A
1 A
20 Ω
200 Ω
19.99
199.9
10 mΩ
100 mΩ
10 mA
10 mA
ranges:
FILE LOCATION: E.F. Disk - Manuals - Cal. Man. - Model 5600.DOC
Case: Fiberglass-charged polycarbonate, with
handle and removable lid, watertight to IP534
(lid closed), IP520 (lid open).
Color: Safety yellow case with gray faceplate
Operating Temperature:
14° to 131°F (-10° to 55°C)
Storage Temperature:
- 4° to 131°F (-20°C to 55°C)
Dimensions:
15.35 x 10.24 x 9.84" (390 x 260 x 250 mm)
Weight: 17 lbs 10 oz (8 kg approx.) without
leads
Safety Specifications
Safety Standards:
Electrical:
IEC 348, Class II
IEC 38, Supply voltage
Electromagnetic:
IEC 801-2, Electrostatic discharge
IEC 801-3, Electromagnetic Fields
IEC 801-5, Electrical shocks
Mechanical/Environmental:
IEC 529, Protection Index (IP)
IEC 68.2.6, Vibrations
IEC 68.2.27, Shocks
IEC 68.2.29, Shake
IEC 68.2.31, Knock over
IEC 68.2.32, Drop test
Case Material:
UL 94
Dielectric Test:
IEC 348, 1500 V 60 Hz (input terminals)
3000 V 60 Hz (power supply)
Overload Input Protection:
380 V rms
Fuses:
Two high interrupting capacity, 8 A, 380 V,
1-1/4" x 1/4"
6