Directions for use
Immunity test
Voltage Dips/Dropout
IEC 61000-4-11
Power frequency 50/60Hz
Magnetic field
IEC 61000-4-8
Conducted RF IEC
61000-4-6
Radiated RF IEC 61000-4-3 3 V, 0.15 MHz – 80 MHz
Compliance level
IEC 60601
Test level
0 % UT; 0.5 cycle At 0°,
0 % UT; 0.5 cycle At 0°,
45°, 90°, 135°, 180°, 225°,
45°, 90°, 135°, 180°, 225°,
270° and 315° % UT; 1
270° and 315° % UT; 1
cycle
cycle
70 % UT; 25/30 cycles for
70 % UT; 25/30 cycles for
50 Hz and 60Hz,
50 Hz and 60Hz,
respectively
respectively
Single phase: at 0°
Single phase: at 0°
0 % UT; 250/300 cycle for
0 % UT; 250/300 cycle for
50 Hz and 60 Hz,
50 Hz and 60 Hz,
respectively
respectively
Single phase: at 0°
Single phase: at 0°
30 A/m, 50 Hz or 60 Hz
30 A/m, 50 Hz or 60 Hz
3 V, 0.15 MHz – 80 MHz
3 V, 0.15 MHz – 80 MHz
6 V in ISM radio bands
6 V in ISM radio bands
between 0.15 MHz and 80
between 0.15 MHz and 80
MHz
MHz
80 % AM at 1 kHz
80 % AM at 1 kHz
3 V, 0.15 MHz – 80 MHz
6 V in ISM radio bands
6 V in ISM radio bands
between 0.15 MHz and 80
between 0.15 MHz and 80
MHz
MHz
80 % AM at 1 kHz
80 % AM at 1 kHz
Standards and compliance 49
Electromagnetic
environment–guidance
Mains power quality should be that
of a typical home, commercial, or
hospital environment. If the user of
the RETeval-DR requires continued
operation during power mains
interruptions, it is recommended
that the RETeval-DR be powered
from an uninterruptible power
supply or battery.
Power frequency magnetic fields
should be that of a typical home,
commercial, or hospital
environment.
Portable and mobile
communications equipment should
be separated from the RETeval-DR
device by no less than the distances
calculated/listed below:
D= (1.17) √P, 150kHz to 80MHz
D= (1.17) √P, 80 to 800 MHz
D= (2.33) √P, 800 MHz to 2.5 GHz
where P is the max power in watts
and D is the recommended
separation distance in meters. Field
strengths from fixed transmitters, as
determined by an electromagnetic
site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the
vicinity of equipment containing a
transmitter.
Portable and mobile
communications equipment should
be separated from the RETeval-DR
device by no less than the distances
calculated/listed below:
D= (1.17) √P, 150kHz to 80MHz
D= (1.17) √P, 80 to 800 MHz
D= (2.33) √P, 800 MHz to 2.5 GHz
where P is the max power in watts
and D is the recommended
separation distance in meters. Field
strengths from fixed transmitters, as
determined by an electromagnetic
site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the