Summary of Contents for NSK FX65

  • Page 2 Chuck Open Open Chuck Close Close...
  • Page 4: User And Intended Use

    • Do not attempt to disassemble the product nor tamper with the mechanism except as recommend by NSK in this Operation Manual. • Do not allow any impact on to the product. Do not drop the product.
  • Page 5 3-2 Disconnection • Should the product function abnormally, cease operation Hold the motor and the handpiece separately and disconnect with immediately and contact your Authorized NSK Dealer. care. • Do not use high acid water or sterilizing solutions to wipe, immerse or clean the handpiece.
  • Page 6: Check Before Treatment

    • Entry of debris into the chuck, via the bur shank, could cause bur Authorized NSK Dealer. rotation slip and also prevent the bur from being securely located in the chuck.
  • Page 7: Maintenance

    6-2 Lubrication 6. Maintenance NSK PANA SPRAY Plus After each patient maintain the product as follows. Apply NSK PANA SPRAY Plus every time after each use and/or before autoclaving. 6-1 Cleaning (Handpiece) 1) Remove the bur from the handpiece. 1) Remove dirt and debris from the product. Do not use a wire brush.
  • Page 8 NSK Care3 Plus CAUTION When using NSK Care3 Plus automatic handpiece cleaning and lubrication system, refer to the Care3 System instructions. • Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and damage 6-3 Sterilization to the product from chemical residue on other instruments.
  • Page 9: Periodical Maintenance Checks

    NSK products are warranted against manufacturing errors and defects Authorized NSK Dealer. in materials. NSK reserves the right to analyze and determine the cause of any problem. Warranty is voided should the product be not used correctly or for the intended purpose or has been tampered with by unqualified personnel or has had non NSK parts installed.
  • Page 10: Disposing Product

    11. Disposing product In order to avoid the health risks of operators handling the disposal of medical equipment, as well as the risks of environmental contamination caused thereof, a surgeon or a dentist is required to confirm the equipment is sterile. Ask specialist firms who are licensed to dispose of specially controlled industrial wastes, to dispose the product for you.
  • Page 11: Specifications

    12. Specifications...