Alarm Messages - Honeywell Envitec MySignO Operating Manual

Oxygen measuring device
Table of Contents

Advertisement

Oxygen measuring device MySign
7

Alarm messages

Visual signals
+
+
+
lower alarm limit
+
lower alarm limit
+
upper alarm limit
Description of the priorities
Low level priority warning
Medium level priority warning
High level priority warning
40
®
O
Audio signals Description
Steady
---
yellow
Flashing
3 x
every
red
20 seconds
Steady
3 x
every
yellow
20 seconds
Flashing
5 x
every
red
10 seconds
Flashing
3 x
every
yellow
20 seconds
Flashing
3 x
every
yellow
20 seconds
Battery charge too low
The battery must be charged
as soon as possible
Battery charge is critically low
The battery must be charged
immediately
No sensor connected
Check the sensor cable and
connector. If necessary,
replace the sensor or cable.
Oxygen concentration is
less than 18%
Check the patient's oxygen
supply and the unit's
calibration as well as the
®
MySign
O sensor.
The oxygen concentration
is too low
Check the patient's oxygen
supply and the unit's
calibration as well as the
®
MySign
O sensor.
The oxygen concentration
is too high
Check the patient's oxygen
supply and the unit's
calibration as well as the
®
MySign
O sensor.
Doc no.: 065-07-1001825_MySignO_GB-0.pdf / 12.11