Oxygen measuring device MySign
Autoset
Automatically sets the upper and lower limit to +/- 3% of the currently shown measured value, e.g.:
Currently measured value
Upper limit
Lower limit
Manual
When configuring the alarm limits manually, the corresponding value must be selected and
edited using the selection buttons once it starts flashing. Once it has been set, confirm the set
value by pressing the right soft key. Press the Home button to return to the measuring mode.
Caution! Danger!
The oxygen concentration of the breathing air supply must not be lower than 18%.
4.8.2

Alarm volume

This function can be used to individually adjust the volume for the audio alarms.
26
24% O
27 %
21 %
®
O
2
Select "Alarm volume" from the "Alarm setting"
menu.
Use the selection keys to adjust the alarm
volume.
Then confirm by pressing the right function key.
Doc no.: 065-07-1001825_MySignO_GB-0.pdf / 12.11