Keysight Technologies E8257D Installation Note page 12

Psg signal generators, add option 1eh improved low band harmonics from 10 mhz to 3.2 ghz, kit
Hide thumbs Also See for E8257D:
Table of Contents

Advertisement

Add Option 1EH (Improved Low Band Harmonics from 10 MHz to 3.2 GHz) Kit Part Number:
E8251-60382
10.Using the needle-nose pliers, connect a cable retainer top to W118.
11.Place a cable retainer base on W76 (E8251-20590).
12.Insert W76 through the A31 Motherboard, at the A38 Lowband Switch Filter (OUT) opening.
13.Using the needle-nose pliers, connect a cable retainer top to W76.
14.Using the 5/16" open-ended wrench, connect the other end of W76 to the
A12 Pulse Modulator.
15.Position the instrument with the RF deck on top and the front panel facing you.
16.Install the A38 Lowband Switch Filter (Option 1EH) assembly.
Figure 5
Placement of W120
W120
12
Installation Note E8251-90382