Ranges
You can select the lower or higher range for CW mode too. Power can be
edited in either of the two ranges. Low range will supply higher accuracy and
better resolution when you set lower power. Any value you set is outside the
maximum value of the LOW RANGE, you should select HIGH RANGE.
Immediate power level
You could set the power level via front panel. If the load is in CW mode, the new
setting power level immediately changes the input at a rate determined by the
slew rate. If the load is not in CW mode, the setting power level will be saved for
use until switch to CR mode.
Transient power level
Set A/B transient power level on front panel or by remote operation, the load
can continuously toggle between the two levels when transient operation is
turned on.

4.3.5 Operations

The following will show you how to set operation mode on front panel. Take CC
mode as an example:
1.
Power on the electronic load
2.
Press
channel 1
CH01 CC OFF
Vdc=0.0000V
Adc=0.0000A
Wdc=0.00W
3.
Press
enter the mode selection menu <Mode>, press
mode CC/CV/CR/CW,such as CC,press
CH01
Mode=CC
RANGE=HIGH
Iset=9.000A
4.
Press
key to enter the mode selection menu <Mode>, press
such as <low range>, press
CH01
Fig4-4
▲ ▲
▼ ▼
or
keys to select the channel to be edited, such as
Setup
Setup
key to enter the channel setup menu, press
▼ ▼
key to enter range selection menu <Range>, press
Enter
Copyright © Itech Electronic Co., Ltd.
Operations introduction
▼ ▼
key to select the
Enter
key to confirm
▼ ▼
key to confirm.
Enter
key to
Enter
key to select,
33