  • ENGLISH, page 21
Checking that the contents of the calibration counter are
correct
This seca scale is calibrated. Calibration may be performed only by authorised centres.
To ensure this, the scale is fitted with a calibration counter which records any change to
data relevant to calibration.The calibration counter stamp can be found on the display
housing.
If you want to check whether the scale is properly calibrated, proceed as follows:
– If necessary, switch off the scale.
– Keep any key depressed and start the
scale.
The current contents of the calibration
counter is flashing on the display for a
few seconds.
– Compare the output contents of the
calibration counter with the number
shown on the calibration counter
sticker.
For valid calibration, both numbers must
match. If sticker and calibration counter
do not match, then re-calibration is re-
quired. Contact your service agent or
seca Customer Service.
Correct weighing
– Press the Start key.
se(A
8. 8 . 8 . 8 . 8 .
,
and
utively in the display. The scale is then
automatically set to zero and ready to
use.
The plug symbol
play.
The scale remains in permanent opera-
tion until it is switched off.
– Get the patient into bed.
– Read patient weight off the digital dis-
play.
If a load greater than 250 kg is placed
St0P
on the scale,
play.
– To switch off the scale, press the Start
key again.
Model 985
0. 0
appear consec-
appears in the dis-
appears in the dis-
Both numbers must match.