Setting Up the Module

1 Grounding Screw
SCXI-1121 Calibration Procedure
Complete the following steps to set up the SCXI-1121 module for
verification:
1.
Remove the grounding screw from the module.
2.
Remove the cover on the module to access the potentiometers.
Refer to Figure 1 for an illustration of steps 1 and 2.
Figure 1. Grounding Screw and Cover Removal
3.
Remove the side plate of the SCXI chassis.
4.
Install the module into slot 4 of the SCXI chassis.
1
2
2 Module Cover
4
ni.com