Follow these guidelines to optimize the connections and environment
during calibration:
Keep connections to the SCXI-1121 module short. Long cables and
wires act as antennae, picking up extra noise and thermal offsets that
can affect measurements.
Use shielded copper wire for all cable connections to the device.
Use twisted-pair wire to eliminate noise and thermal offsets.
Maintain the temperature between 18–28 °C.
Keep relative humidity below 80%.
Allow a warm-up time of at least 15 minutes for the SCXI-1121
module to ensure the measurement circuitry is at a stable operating
temperature.
The calibration procedure for the SCXI-1121 module consists of the
following steps:
1.
Setting up the module for testing.
2.
Verifying the existing operation of the module to determine whether it
is operating within its specifications.
3.
Adjusting the module with respect to a known voltage source.
4.
Verifying that the module is operating within its specifications after
adjustments.