4-3-3. Result evaluation
• Go/No Go test is carried out by comparing data from the
current measuring mode (current value, maximum value,
minimum value, or peak-to-peak value) to the comparator
upper and lower limits.
• This result is displayed on the front panel and output from
the terminals (See P60 "6. Terminals I/O".).
Result
Display
High
>
Data > upper limit
Upper limit > data > lower limit
Go
\
Low
?
Condition
=
Lower limit > data
=
4-3-4. Peak-hold function
• Stores the maximum, minimum, and peak-to-peak
(maximum–minimum) values of the measured values.
• The above mentioned measuring mode is set using the
keys on the front panel.
• The device starts storing values on receiving a start
signal input from pin 0 of the terminals, or when the
reset key has been pressed.
Operation
Start on "L" (ON) signal
Starts storing from the current value.
from pin 0 of the terminals
Reset key is pressed
Starts storing from zero.
When a preset value is set the
device starts storing from the preset
value.
Result
51