Main Display Area

Depending on the selection made using the tab controls, the content of the main display area is changed to either the
profile graph, the pulse data graph or the stimulation data table.

Profile Graph

The profile graph displays all stimulation results of the current session, indicating the result, the depth-of-block zone and
the relative time from the start of the session. Only Auto, Train of Four (TOF) and Post Tetanus Count (PTC) modes are
supported. Unsupported modes' results will be shown as a grayed-out dot on the graph. The graph is subdivided into the
following six depth of block zones:
Recovered – Train of four (TOF) ratios over 90% are classified as recovered.
Minimal – TOF ratios between 40% and 90% are classified as a minimal block.
Shallow – A shallow block corresponds to TOF ratios between 10% and 40%.
Moderate – TOF counts between one and three are classified as a moderate block. TOF counts of four with a ratio
less than 10% technically constitutes the boundary between Shallow and Moderate.
Deep – A TOF count of zero, or a PTC of one or more are classified as deep.
Profound – If a PTC of zero has been observed, the depth-of-block is classified as profound.
Please note that occasionally the communication may be interrupted due to either electromagnetic noise, user input or
communication malfunction. Any failed stimulation sequence will be grayed out for this stimulation on the graph and the
zone will be set to Failed.